Kunming Box type Resistance Furnace (Precision Type)Specification parameters:
model |
AS-1400 |
Temperature parameters |
1400℃ |
Voltage AC |
220V/380V |
rated temperature |
1400℃ |
Long term working temperature |
1350℃ |
Uniformity of temperature field inside the furnace |
±5℃ (depending on the size of the heating chamber) |
Temperature measuring components and temperature measuring range |
Platinum rhodium platinum S temperature measurement range 0-1700 ℃ |
Program curve segment |
30 paragraphs |
heating rate |
1 ℃/h-40 ℃/min, recommended 20 ℃/min |
Heating element |
Elema |
Thermal insulation material |
Imported high-purity alumina fiberboard |
Cooling method |
Double layered furnace shell, air-cooled |
Furnace body temperature |
≤ 50 degrees |
Warranty scope and duration |
The electric furnace comes with a one-year free warranty, and the heating element is not covered by the warranty (natural damage within three months can be replaced free of charge) |
Customer selection |
1. Crucible, box bowl, corundum furnace pad/silicon carbide pad 2. RS485 communication, computer control software and hardware 3. Touch screen controlled temperature controller 4. Paperless recorder 5. Increase exhaust ports 6. Consumables: Heating elements, temperature measuring components 7. Increase observation ports and endoscopes |
Furnace specifications:
specifications |
Furnace dimensions (depth, width, height mm) |
voltage |
Power kw |
PID instrumentControl accuracy |
A |
100x100x100 |
220 |
1 |
±1 |
B |
150X100X100 |
220 |
1.5 |
±1 |
C |
200X150X150 |
220 |
2 |
±1 |
D |
300X200X120 |
220 |
4 |
±1 |
E |
200×200×200 |
220 |
6 |
±1 |
F |
300×200×200 |
220 |
8 |
±1 |
G |
300X250X250 |
220/380 |
8 |
±1 |
H |
300×300×300 |
220/380 |
10 |
±1 |
I |
400X250X160 |
220/380 |
8 |
±1 |
J |
400×300×300 |
220/380 |
12 |
±1 |
K |
500×300×200 |
220/380 |
15 |
±1 |
L |
500×300×300 |
380 |
18 |
±1 |
M |
500X400X400 |
380 |
20 |
±1 |
N |
500×500×500 |
380 |
25 |
±1 |
U |
800×500×500 |
380 |
40 |
±1 |
V |
1200×500×500 |
380 |
85 |
±1 |
W |
1200×800×800 |
380 |
110 |
±1 |
Kunming Box type Resistance Furnace (Precision Type)Special furnace dimensions can be customized according to customer requirements!
Reasons and solutions for temperature control faults in ceramic muffle furnaces
Temperature control failure
reason:
1. The temperature gauge has been disturbed;
2. The thermocouple signal has been disturbed;
3. Poor contact of thermocouple;
4. Compensate for wire short circuits or road robberies;
5. The thermocouple is too close to the electric heating component;
6. Temperature control circuit malfunction;
7. The thermocouple has been corroded and its electrical characteristics have undergone changes;
8. Thermocouples are grounded, and the insulation of the shielding tube is raised;
Solution:
1. Take measures to eliminate sources of disturbance;
2. Increase barriers for signal lines;
3. Tighten the thermocouple reporting terminal;
4. Review and make up for the wires;
5. Adjust the interval between the two;
6. Repair the temperature control circuit;
7. Replace the thermocouple;
8. Review the insulation condition of thermocouples;