Karl Fischer titration moisture analyzer [Coulombic titration method] MKC-710D

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Karl Fischer titration moisture analyzer MKC-710D supports data integrity, including access restrictions, permission management, audit trails, electronic records/signatures, backup/restore/archive, data capacity, security settings, LIMS compatibility, and other functions. Meets GMP, GLP, and FDA 21 CFR Part 11 specifications.

Product Details


Karl Fischer titration moisture analyzer [Coulombic titration method] MKC-710D

Support data integrity, including access restrictions, permission management, audit trails, electronic records/signatures, backup/restore/archive, data capacity, security settings, LIMS compatibility, and other functions.
Using a large color touch screen with wireless connection, it can control two measuring units.
MKC-710M/S can be upgraded to the MKC-70D model through upgrade settings.

Karl Fischer titration moisture analyzer [Coulombic titration method] MKC-710D Main features:

1. Adopting a large 8.4-inch color LCD touch screen for operation and control.
2. The touch screen is more secure and reduces the risk through wireless Bluetooth.
3. The measurement results can be stored on a USB drive and generate a PDF experimental report.
4. Real time display of moisture content and curves. And it has energy-saving and power-saving functions.
5. The reagent reaction ability of the anolyte and catholyte can be set separately.
6. The electrolysis rate is 2.6mg (2600 μ g) H2O/min, and the measurement speed is fast.

Karl Fischer titration moisture analyzer [Coulombic titration method] MKC-710D technical parameters:

Name and Model

Karl Fischer moisture analyzer MKC-710D

measurement range

Moisture content: 1 μ g-300mgH2O

Moisture content: 1ppm~99% H2O

Bromine index: 8 μ g-300mgH2O

Measurement accuracy

Relative standard deviation less than 0.3% (n=10)

resolution

0.1μg

Endpoint detection

Current pulse time control

Terminal judgment

Drift value stabilization time control or set measurement time

Reagent demand

Anode solution: 100mL (150mL)

Cathodic liquid: 5mL

Other functions

1) Drift value automatic adjustment

2) Automatic start measurement

3) Can store 10 sets of blank values

4) Optional automatic or manual suction and discharge device

Button operation

Touch screen direct operation

display

1) 8.4-inch color LCD screen, resolution: 800 × 600

2) Chinese/English/Japanese operating interface

3) Dual channel simultaneous display

calculate

Concentration calculation, statistical calculation, automatic input of blank value and titration degree

data storage

120 measurement methods, 5000 sample results, 100000 operation records

Data integrity function

Access restriction: Authentication through personnel ID and password

Permission management: Assign each operator to a specific permission group

Audit Tracking: Automatically Record Historical Operation Information

Electronic Records/Electronic Signature: Electronic approval and signature based on workflow

Backup/Restore/Archive: Automatically backup data at any interval

External output/input

RS-232C×4,USB×1,SS-BUS×1,LAN×1