JPL-TD-2 automatic secondary thermal desorption desorption instrument

NegotiableUpdate on 03/17
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The JPL-TD-2 automatic thermal desorption desorption instrument is an automatic secondary thermal desorption instrument independently developed and launched by Shandong Jinpu, which has a fully automated design, touch screen display, and more convenient operation. The gas circuit adopts a combination of electric six way valves and solenoid valves, and uses a touch screen for easy editing, saving, and calling of the desorption, injection, and blowback processes of the adsorption tube. It can clearly display the action status of each part during the sample extraction process, and users can easily understand the operation process of the instrument. The heating temperature of adsorption tubes, six way valves, pipelines, etc. can be independently set, and synchronous signals can be output during injection to simultaneously start chromatography and workstations.
Product Details

The product meets the requirements for thermal desorption analyzer in standards such as GB50325-2010 Code for Indoor Environmental Pollution Control of Civil Building Engineering and HJ 583-2010 Determination of Benzene Compounds in Environmental Air Solid Adsorption/Thermal Desorption Gas Chromatography.

2、 Instrument features and main functions

1. The gas circuit adopts a combination of electric six way valves and solenoid valves to automatically achieve desorption, injection, and blowback cleaning processes

2. Adopting a touch screen for convenient editing, saving, and calling methods

3. The heating temperature of the adsorption tube, pipeline, and six way valve can be independently set,

4. The machine comes with the function of simulating and sampling standard samples, which makes it more convenient to create working curves through a thermal desorption instrument;

5. Through time programming, automatic functions such as desorption, blowdown adsorption, injection, and blowback cleaning (activation) can be achieved;

6. Using Simple Lock locking sealing technology to achieve simple and reliable fixation and sealing of sampling tubes, efficient work can be achieved by fixing and disassembling the sampling tubes in 2 seconds

7. Adopting semiconductor refrigeration, high cooling efficiency, no need for external refrigeration equipment, saving usage costs

7. Using standard adsorption tubes (89mm long, 6.4mm outer diameter), the adsorption tubes can withstand a temperature of 420 ℃ and are sealed with PTFE caps to ensure longer storage time of the samples;

8. The injection needle is easy to replace and can be connected to all models of GC injection ports at home and abroad

3、 Main technical parameters of the instrument

1. Temperature control range of the pipette: room temperature -400 ℃, accuracy: 1 ℃;

2. Temperature control range of valve injection system: room temperature -220 ℃, accuracy: 1 ℃;

3. Temperature control range of sample transfer pipeline: room temperature -260 ℃, accuracy: 1 ℃;

4. Temperature range for secondary analysis: -35 ℃ -400 ℃, heating rate for secondary flash evaporation:>3000 ℃/min

5. Desorption recovery rate:>98% (depending on the composition);

6. Backblowing cleaning flow rate: 0-600ml/min (continuously adjustable);

7. Simulated sampling flow rate: 100ml/min;

configuration

1. One host;
2. Two American standard Tenax TA thermal analysis tubes;
3. Simple Lock locked sealed injection system;
4. One gas injection port adapter bracket (different brands have different configurations, and the brand needs to be specified before procurement);
5. One set of standard sample simulation sampling device;
6. 2 meters of PTFE gas pipeline;
7. One stainless steel tee;

热解析仪厂家常见问题常见问题常见问题荣誉证书