The cold and hot shock test chamber can be divided into two chamber and three chamber types, with the difference being the testing method and internal structure. It is a testing equipment used in the metal, plastic, rubber, electronics and other material industries. The manual of the cold and hot shock test chamber is used to test the material structure or composite materials, and to determine the degree of tolerance they can withstand under continuous environments of high and low temperatures in an instant, and to detect the chemical changes or physical damage caused by thermal expansion and contraction of the sample.

Product Details

Technical performance of cold and hot shock test chamber:

Instructions for Cold and Hot Shock Test ChamberThe inner box material is made of SUS # 304 stainless steel plate, and the outer box material is made of SEEC steel plate with electrostatic spraying or stainless steel drawing plate, which increases the appearance and cleanliness of the machine. The dead corner circular arc design of the box body is beautiful and elegant.

The insulation layer of the box adopts high-strength PU and insulation cotton to achieve excellent insulation effect.

The box is equipped with a test hole with a diameter of 50mm, which can be used to test components for external power load wiring.

Refrigeration and Heating Control System:

1) Refrigerator: High performance refrigeration compressor imported from Europe and America (French Taikang fully enclosed compressor or German Gulun semi enclosed compressor)

Cooling method: air-cooled binary duplex system

Refrigerant: R404A, R23

2) Heater: finned heat dissipation tube shaped nickel chromium alloy U-shaped high-efficiency heater

3) Control system: * Microcomputer large color LCD display touch controller, Chinese and English display control system. Equipped with RS-232 communication interface and connection software, it can design programs, collect test data and records, call program execution on the computer screen, and achieve remote control.

1. High temperature storage range:+60 ℃ -+200 ℃

2. Low temperature storage range: -10 ℃~-55 ℃ (Type A), -10 ℃~-65 ℃ (Type B), -10 ℃~-75 ℃ (Type C)

3. Test scope:

High temperature:+60 ℃~+150 ℃

Low temperature: -10 ℃~-40 ℃ (Type A); -10 ℃~-55 ℃ (Type B); -10~-60 ℃ (Type C)

4. Heating time:+60 ℃~+200 ℃, approximately 25 minutes required

5. Cooling time:+20 ℃~-55 ℃ less than 60 minutes; +20 ℃~-65 ℃ less than 75min; +20 ℃~-75 ℃ less than 90 minutes;

6. Wind door switching time: less than or equal to 10 seconds

7. Temperature recovery time: 5 minutes

8. Temperature fluctuation: ± 0.5 ℃

9. Temperature deviation: ± 2 ℃

Model: BY-260D

① Specification (Capacity): 63 liters

Test area size: 400 * 350 * 350 (W * D * Hmm)

Dimensions (approximately): 1500 * 1450 * 1900 (W * D * Hmm)

② Specification (Capacity): 80 liters

Test area size: 500 * 400 * 400 (W * D * Hmm)

Dimensions (approximately): 1600 * 1500 * 2000 (W * D * Hmm)

③ Specification (Capacity): 150 liters

Test area size: 600 * 500 * 500 (W * D * Hmm)

Dimensions (approximately): 1700 * 1600 * 2000 (W * D * Hmm)

④ Specification (Capacity): 225 liters

Test area size: 500 * 600 * 750 (W * D * Hmm)

Dimensions (approximately): 1800 * 1800 * 2000 (W * D * Hmm)

⑤ Specification (Capacity): 500 liters

Test area size: 800 * 800 * 800 (W * D * Hmm)

Dimensions (approximately): 2600 * 1900 * 2000 (W * D * Hmm)

⑥ Specification (Capacity): 1000 liters

Test area size: 1000 * 1000 * 1000 (W * D * Hmm)

Dimensions (approximately): 2800 * 2000 * 2200 (W * D * Hmm)

Free home delivery, installation and debugging operation introduction, and machine operation technical training. (Until the demand side employees operate independently and are satisfied).

This product complies with standards such as GB/T2423.1.2 GB10592-89 GJB150.

Instructions for Cold and Hot Shock Test ChamberThe equipment adopts a modular structure, consisting of a high-temperature test chamber located at the upper part, a low-temperature test box located at the lower part, a refrigeration unit cabinet located at the rear, and an electrical control cabinet (system) located on the left rear panel This method has a small footprint, compact structure, and beautiful appearance. The refrigeration unit is placed in an independent unit box to reduce the impact of vibration and noise during the operation of the refrigeration unit on the test box.

*Temperature and humidity controller, with simple operation, easy learning, high control accuracy, * PID mode, can save energy by more than 10%.

It can independently set three different conditions: high temperature, low temperature, and thermal shock, and has the function of a high and low temperature testing machine.

The number of cycles and defrosting times can be set for automatic (manual) defrosting The air outlet is detected and controlled by the return air outlet sensor, and the switching time of the air door mechanism is completed within 10 seconds.

Equipped with fully automatic, high-precision system circuits, any component action, * with P.L.C locking processing, all using P.I.D automatic calculation control.

The temperature recovery time is completed within 5 minutes.

When an abnormal situation occurs during operation, the screen will automatically display the cause of the fault, and an emergency stop device will be installed when unstable input power is detected.

Optional USB interface for convenient storage, with standard RS-232 or RS-485 computer communication interfaces available. Implement remote control.

The frequency of application of cold and hot shock test chambers in verifying and improving the environmental adaptability of equipment is second only to vibration and high and low temperature tests. It is used to assess the adaptability of products to rapid changes in ambient temperature, and is a test in equipment design and batch production stages. In some cases, it can also be used for environmental stress screening tests