Infrared analyzer

NegotiableUpdate on 03/13
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The selective absorption of infrared radiation by gas molecules is the design basis of infrared gas analyzers
Product Details
Product Features:
Specially designed gas chamber absorber, narrowband filter with strong resistance and high detection accuracy
◆ High stability infrared light source, receiver temperature control, good instrument stability
Measurement chamber mirror with gold plating, corrosion-resistant
◆ Microcomputer intelligent modular design, button operation, LCD large screen blue screen display
Isolation output signal 0/4-20mA Dc, optional communication interface
◆ Alarm output (upper and lower limit extreme value alarm, self-test fault alarm)
◆ Timer automatic storage function, allowing users to view stored data and curves at any time
Technical Specifications:
Measurement principle: Non dispersive infrared method: NOX, SO2、CO、CO2、 CH4
Measurement components: NOx, SO2、CO、CO2、CH4It can be measured that any one or two components O2 can be added to each component
Measurement range: The measuring range can be designed according to requirements for the smallest range and the largest range
NOx 0~1000ppm 0~20%
SO20~500ppm 0~100%
CO 0~500ppm 0~100%
CO20~100ppm 0~100%
CH4 0~200ppm 0~100%
Repeatability: ≤± 1% F.S
Zero drift: ≤± 1.5% F.S./7d
Full range drift: ≤± 2.0% F.S./7d
Linearity: ≤± 2.0% F.S
Response time: T90≤ 60 seconds (when the sample flow rate is 0.5L/min)
Output signal: Analog output: 4-20uA
Maximum load: 750 Ω digital communication RS485 (optional)
Environmental conditions: 5 ℃~40 ℃, not exceeding 90% RH
Power supply voltage: AC220V ± 10% 50Hz
Dimensions: 4U panel 440mm × 160mm × 400mm or 440mm × 160mm × 500mm (W × H × D)
Condition of the measured gas: temperature 0~+50 ℃
Flow rate: 0.5 l/min ± 0.2 1/min
Dust: particles not exceeding 1 μ m, not exceeding 200 μ g/Nm3
Aerosol: None
Pressure: not exceeding 10kpa (the outlet end of the measured gas should be at atmospheric pressure)
Moisture: not exceeding 5 ℃ saturated (no condensation)
Ordering Notice (Please specify when placing an order)
Instrument measurement range
Tested gas pressure: positive pressure, slight pressure, or negative pressure
Main components, impurities, and sulfides of the tested gas