E-mail
hengqin17@126.com
-
Phone
13361860330 / 15021277021
-
Address
Room 302, No. 28, Sitangba Village, Baoshan District, Shanghai
Shanghai Hengqin Instrument Equipment Co., Ltd
hengqin17@126.com
13361860330 / 15021277021
Room 302, No. 28, Sitangba Village, Baoshan District, Shanghai
| Extremely fast drying The circular halogen lamp and gold-plated radiator ensure reproducibility and rapid heating of the sample. Reliable measurement results can be obtained in just a few minutes. | |
| Accuracy and repeatability The outstanding performance of HR83 is based on Mettler Toledo's analytical weighing technology. Obtain accurate results with a resolution of up to 0.001% MC and the highest measurement safety. | |
| Compliant with SOP and relevant regulations HR83 is an ideal choice for regulatory environments, providing product features that comply with regulatory requirements, including password protection, sample identification, and traceable instrument calibration reports. The intelligent solution comes from Mettler Toledo HR83. | |
| AutoMet automatic measurement method development function The AutoMet function can help you easily obtain measurement results that match the reference values. If you need to develop a measurement method, you only need to enter the target value, and HR83 will help you complete the remaining work. By using the AutoMet function, the development time of measurement methods can be shortened by about 50%. | |
|
IPac Moisture IPac Moisture is an effective certification tool to ensure quality and regulatory compliance:
|
| temperature range | 40 - 200°C (104 - 392 °F) |
| Readability of moisture content | 0.01% / 0.001% |
| temperature regulation | 1 °C |
| Small sample weighing | 0.1 g |
| Large sample weighing | 81 g |
| Repeatability (s) (2g sample) | 0.05 % |
| Repeatability (s) (10g sample) | 0.01 % |
| Drying program | 4 types (standard, fast, slow, step-by-step drying) |
| Cut off standards | Automatic shutdown (level 5 and free shutdown), manual, time control |
| printer | Built in printer options |
| Weight, ready for measurement | 7.7 kg (including built-in printer) |
| monitor | LCD backlit LCD display screen |
| interface | LocalCan/RS232 interface (optional) |
| language | English, French, German, Italian, Russian, Spanish |
| size | 360 x 340 x 110 mm (DxH) |
| unit | Moisture content%; Solid content%; Quality; ATRO moisture content%; ATRO solid content% |
| readability | 1 mg / 0.1 mg |
| Method Development | AutoMet and testing functions |
| Sample Identification | 4 comment lines, can be used in conjunction with the barcode reader option |
| Balance test | Test result record: 0.1 g to 80 g (variable verification) |
| Heating module testing | Three temperature testing points, including the temperature at the measuring point |
| Additional features | Storage of 40 methods, graphical weighing assistant, free factor, measurement method link and standby temperature, statistical and logging functions, password protection, GLP/GMP compliant print output |