Welcome Customer !

Membership

Help

Shanghai Hanyao Instrument Equipment Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products
Product Categories

Shanghai Hanyao Instrument Equipment Co., Ltd

  • E-mail

    sales@shhanking.com

  • Phone

    13916306041

  • Address

    Room 2908-2909, Greenland Science and Technology Island Plaza, No. 2, Lane 58, Shenjian East Road, Minhang District, Shanghai

Contact Now

German binder BF series forced convection incubator

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Product Details

German binder BF series forced convection incubator

BF Series | Avantgarde. Line Forced Convection Incubator

The BF series is suitable for laboratories that require batch cultivation, and the new generation APT.lineTM technology provides uniform temperature distribution for the enclosure. The controller has a timing function, which can easily adjust the fan, temperature, and forced convection situation.

Product performance characteristics:

•APT. LineTM preheating chamber technology ensures uniform temperature distribution and accuracy

•LCD display

•Mechanical controlled ventilation opening

•Tempered glass inner door

•2 chrome plated shelves

•Products below 115L can be stacked and placed

•Independent adjustable Class3.1 temperature safety device (compliant with DIN12880 standard) with indicator light alarm function

•The door handle conforms to the principles of ergonomics

•USB port

Common optional accessories

•Test hole with silicone plug

•Layer frame, chrome plated or stainless steel layer frame

•door lock

•Ethernet data port

BF series biochemical incubator

The BF series forced convection incubator is very suitable for large-scale sample cultivation and situations requiring short temperature recovery time. Its outstanding feature is its high temperature control accuracy, providing constant and reproducible cultivation conditions.

1) Forced convection electronic control APT. LineTM cavity preheating technology;

2) Temperature range: 5 ℃ to 100 ℃ above ambient temperature;

3) Digital temperature setting with an accuracy of 0.1 ℃;

4) MS controller, with several timer functions;

5) The timer function of the program controller includes delayed activation, delayed deactivation, and temperature dependent delayed deactivation;

6) The heating rate can be adjusted through a program editor;

7) Fan speed adjustable (0 to 99%);

8) Adjust the ventilation volume through the exhaust pipe with a ventilation flap on the back (diameter of 50 millimeters) and the ventilation skateboard in front;

9) Independent adjustable temperature safety device, Class 3.1(DIN 12880), Equipped with a visual alarm;

10) Internal glass door;

11) Boxes of 115L or below can be stacked, with 2 chrome plated shelves;

12) RS 422 interface, used for communication software APT-COMTM data control system; Or use the RS 232/RS 422 interface converter to convert to the printer output port; The interval between printers can be adjusted;

model

BF 53

BF 115

BF 240

BF 400

BF 720

External dimensions (W × H × D) (mm)

634×617×

575

834×702×

645

1034×822×

745

1234×1022×765

1234×1528×865

Internal cavity size (W × H × D) (mm)

400×400×

330

600×480×

400

800×600×

500

1000×800×

500

1000×1200×600

Internal cavity volume (L)

53

115

240

400

720

Number of shelves (standard/)

2 / 5

2 / 5

2 / 7

2 / 10

2 / 15

Load per shelf/total load (kg)

15 / 40

20 / 50

30 / 70

35 / 90

45 / 120

Number of doors/Internal glass doors

1 / 1

1 / 1

2 / 2

2 / 2

2 / 2

Temperature range: ambient temperature+5 to (℃)

100

100

100

100

100

Temperature deviation @ 37 ℃ (± ℃)

0.4

0.3

0.3

0.4

0.4

Temperature deviation @ 50 ℃ (± ℃)

0.7

0.6

0.8

0.9

0.6

Temperature fluctuation range @ 37 ℃ (± ℃)

0.2

0.2

0.2

0.2

0.1

Temperature fluctuation range @ 50 ℃ (± ℃)

0.2

0.2

0.2

0.2

0.2

Heating time to 37 ℃ (minutes)

12

22

12

18

21

Heating time to 50 ℃ (minutes)

20

23

24

26

24

Recovery time (30 seconds after opening the door to 37 ℃) (min)

1

1

1

2

1

Recovery time (30 seconds after opening the door to 50 ℃) (min)

1.5

2

2

4

4

Power supply (± 10% 50/60Hz)

230

230

230

230

230

power (kW)

0.4

0.4

0.68

0.85

1.25

Net weight (kg)

43

64

104

145

180