| Just like the filament in an incandescent lamp, the filament in an ion source will eventually burn out. The following operations can prevent premature damage to the filament: |
| When setting the acquisition parameters, set the solvent delay so that the analyzer is turned off when the solvent peak flows out |
| When the software prompts' Override solvent delay at the beginning of a run ', be sure to select' No ' |
| Higher emission current will reduce the lifespan of the filament |
| If you are controlling MSD on the parameter editing screen, be sure to select "MS Off" before changing any filament parameters |
| Filament Assembly (EI) G3170-60050 G3170-60050 G3170-60050 05971-60140 |
| Filament assembly (CI) G1099-80053 G1099-80053 |
| Micro ion vacuum gauge G3170-80001/Tripole gauge tube for measuring vacuum 0960-0897/Ion gauge controller G3880-80010/Ion gauge tube G3880-80011 |