G3170-60050 MSD Filaments filaments
Just like the filament in an incandescent lamp, the filament in an ion source will eventually burn out. The following operations can prevent premature damage to the filament:
When setting the acquisition parameters, set the solvent delay so that the analyzer is turned off when the solvent peak flows out
When the software prompts "Override solvent delay at the beginning of a run" at the beginning of the run
Be sure to choose 'No' if it's too late
Higher emission current will reduce the lifespan of the filament
If you are controlling MSD on the parameter editing screen, be sure to select "MS Off" before changing any filament parameters
Explanation 5975 Series 5975T Series 5973 Series 5972 Series 5971 Series
Filament Assembly (EI) G3170-60050 G3170-60050 G3170-60050 05971-60140
Filament assembly (CI) G1099-80053 G1099-80053
Micro ion vacuum gauge G3170-80001
Measuring vacuum with triode gauge 0960-0897
Ion gauge controller G3880-80010
Ion gauge G3880-80011