The RS-3 melting point meter uses PID (temperature automatic control) technology to control temperature and is a product of our company. The characteristic of the instrument Z is its dual use of visual measurement and automatic measurement, which combines visual observation measurement and photoelectric detection automatic measurement on one instrument. In this way, the automatic measurement function of photoelectric detection is convenient for users to use; It can also visually measure the melting point of substances to meet the needs of melting point measurement for dark colored samples. So, the instrument measures a wide range of samples and is also very convenient to use. The instrument can measure three samples simultaneously, with a large touch screen display and typing, and has R
Product Details

RS-3 melting point meteradoptPID(Automatic temperature control) technology controls temperature, which is a domestic product of our company.

The biggest feature of the instrument is thatDual use of visual measurement and automatic measurementThat is, combining visual observation measurement and photoelectric detection automatic measurement on one instrument. In this way, the automatic measurement function of photoelectric detection is convenient for users to use; It can also visually measure the melting point of substances to meet the needs of melting point measurement for dark colored samples. So, the instrument measures a wide range of samples and is also very convenient to use.

This instrument can measure three samples simultaneously, with a large touch screen display and typing, and hasRS232Serial port andUSBInterface.

Main technical parameters:


Melting point measurement range: room temperature-320

The smallest temperature digital display value:0.1
Linear heating rate:0.2 /min1/min1.5/min3/min

Indication error: less than200 Within the temperature range: ±0.5
200 ℃-300Within the temperature range: ±0.8

Repeatability of indication: The heating rate is1.0℃/minWhen,0.3

Standard capillary size: outer diameter Φ1.2mmInner diameter Φ1.0mmheight120mm

Sample filling height: ≥3mm

power supplyAC220V±22V,100W , 50Hz

Output communication interface:RS232TheUSB