Haoliang Optoelectronics Company has launched the Cronus series spectral colorimeter, which can repeatedly measure the color of objects, with total light output and optional flicker and start-up behavior within a few milliseconds. So the Cronus series spectral colorimeter is very suitable for applications such as lighting fixtures and displays that require high speed and accuracy for measurement. Moreover, this spectral colorimeter has been developed for industrial applications that require extreme ease of use, minimal calibration, high-speed measurement, and robust and compact design, allowing ordinary personnel to obtain laboratory level measurement results. Cronus Spectral Colorimeter Integrated Machine
Cronus Spectral Colorimeter Integrated Machine
Haoliang Optoelectronics Company has launched the Cronus series spectral colorimeter, which can repeatedly measure the color of objects, with total light output and optional flicker and start-up behavior within a few milliseconds. So the Cronus series spectral colorimeter is very suitable for applications such as lighting fixtures and displays that require high speed and accuracy for measurement. Moreover, this spectral colorimeter has been developed for industrial applications that require extreme ease of use, minimal calibration, high-speed measurement, and robust and compact design, allowing ordinary personnel to obtain laboratory level measurement results.
Main features:
lCombination of spectrometer and colorimeter
lLED、 Lighting fixtures and display measuring equipment
lmeasurementxy,CRI, Luminous flux, spectral distribution, flicker
lBrightness measurement speed up to50000 times per second
lspectral range380-780nm, Suitable for chromaticity measurement
lMultiple optical configurations (integrating sphere, fiber input, lens, cosine calibration)……)
lAutomatic range modulation function
lMechanical shutter
l*Linearity of
lDark current compensation
lCompliant with USBTMC standards, SCPI command set, high-speed devices
Basic Parameters:
Crouns Spectral Measurement System
spectral range |
380-780nm |
Spectral resolution(FWHM) |
2.3nm |
integration time |
2.5ms-20s |
nonlinearity |
<0.5% |
dynamic range |
>150,000 |
accuracy(Y,x,y) |
4%,+/-0.002 |
repetitiveness(Y,x,y) |
<1%,0.0005 |
Crouns colorimeter measurement system
detector |
adoptXYZSilicon photodiode for interference filter |
spectral response |
approachCIE1931Chromaticity matching curve |
dynamic range |
>10,000,000 |
measure speed |
Brightness measurement>50,000time/second |
accuracy(Y,x,y) |
4%,+/-0.002 |
repetitiveness(Y,x,y) |
<0.5%,0.0002 |
System Configuration
Cosine corrector |
Fixed or connected through fiber optic cables25px2Cosine corrector |
integrating sphere |
75mm, 150mm, 250mm diameter integrating spheres are available for selection |
lens system |
10mm and 20mm fixed lenses are available for selection; 5mm, 10mm, 20mm can be selected through fiber optic connection; |
interface |
high-speedUSBTheRS232Ethernet, external trigger interface |
Shutter lifespan |
superior to100Ten thousand times |
Operating Temperature |
10-35℃ |