SteriTransfer™Aseptic transfer warehouse manufacturerUsed for biological decontamination treatment of the outer surface of materials to prevent contamination from entering A/B critical areas from low-level clean areas. It can be used for various types of cleaning that need to be transferred in aseptic production, for biological decontamination treatment of the outer surface of materials, to avoid contamination caused by materials entering A/B key areas from non level areas/low-level clean areas. Can be used for various clean and dry items that need to be transferred in aseptic production, such as packaging materials, instruments, raw material packaging, accessories, environmental monitoring materials, etc.
Aseptic transfer warehouse manufacturerPerformance characteristics:
1. SteriTransfer ™ The sterile transfer chamber uses hydrogen peroxide steam (VHPS) as a biological cleaning agent, which belongs to the low-temperature and atmospheric pressure cleaning process, efficient and hygienic. The hydrogen peroxide gas sterilization method is also included in the "sterilization method" in the pharmacopoeias of the United States, China and other countries;
2. SteriTransfer ™ The sterile transfer chamber is controlled by SIEMENS programmable logic controller (PLC) to automate the operation of various stages of the equipment, ensuring stable and reliable system operation;
During the biological decontamination and exhaust stage, the air sent into the cabin is filtered through an H14 HEP filter to prevent contamination of the material;
4. SteriTransfer ™ The sterile transfer chamber has an independent ventilation and residual removal unit, including a fan, pneumatic butterfly valve, and exhaust pipeline, which can quickly replace the hydrogen peroxide gas in the chamber and prevent it from entering the HAVC system;
5. The feeding and discharging are double door structures with pneumatic sealing, pneumatic locking, and double door interlocking functions in working state;
6. SteriTransfer ™ The sterile transfer chamber is equipped with a trolley type material support and its track system inside, which facilitates material loading and unloading;
7. The system has real-time monitoring functions for temperature, humidity, and pressure inside the cabin, and can be equipped with ports for detecting hydrogen peroxide gas concentration, dust particles, and planktonic bacteria;
8. SteriTransfer ™ The sterile transfer chamber is easy to verify, with complete IQ, OQ, and PQ documents, and can provide on-site verification.
1. Chamber material: SUS 304 stainless steel
Note: If SUS316 material is used, it should be specified in the purchase contract
2. High efficiency filter level: HEPA for air supply: H14 level, HEPA for exhaust: H14 level
3. Alarm: High voltage/low voltage alarm; Open door timeout alarm; Lock door alarm
4. Biological decontamination cycle time:<120min
Note: Verify and determine based on the biological load level and loading method of the material.
5. Biological decontamination efficacy: lg6 (thermophilic fat spore, ATCC12980 or ATCC 7953)
6. VHPS residual level: ≤ 1ppm
Note: The residual level is determined by the process requirements.
7. Cabin cleanliness (particle level): Class B
8. Hydrogen oxide gas decontamination circulation interface: DN32 sanitary interface
9. Exhaust interface: DN100 sanitary interface
10. It has the function of storing and printing biological decontamination cycle parameter data
Technical parameters:
1. Chamber material: SUS 304 stainless steel
Note: If SUS316 material is used, it should be specified in the purchase contract
2. High efficiency filter level: HEPA for air supply: H14 level, HEPA for exhaust: H14 level
3. Alarm: High voltage/low voltage alarm; Open door timeout alarm; Lock door alarm
4. Biological decontamination cycle time:<120min
Note: Verify and determine based on the biological load level and loading method of the material.
5. Biological decontamination efficacy: lg6 (thermophilic fat spore, ATCC12980 or ATCC 7953)
6. VHPS residual level: ≤ 1ppm
Note: The residual level is determined by the process requirements.
7. Cabin cleanliness (particle level): Class B
Typical application: Used for delivering raw materials such as pharmaceuticals and tools into B-level areas

8. Hydrogen peroxide gas decontamination circulation interface: DN32 sanitary interface
9. Exhaust interface: DN100 sanitary interface
10. It has the function of saving and printing biological decontamination cycle parameter data
Typical application: Used for delivering raw materials such as pharmaceuticals and tools into B-level areas