Used for non-destructive contact/non-contact analysis testing and inspection, multifunctional from laboratory to production workshop and on-site.

Product Details

Product Overview

The 5030 "Handheld" Micro AOTF-NIR (Acousto Optic Tunable Filter Near Infrared) Analyzer is a sturdy and durable handheld analytical instrument that utilizes Brimrose's high-performance AOTF-NIR technology. The use of Luminar 5030 provides a powerful new tool for quality control and process troubleshooting by providing fast, multi-component analysis for various applications including diffuse reflectance measurements of powders, particles, liquids, meshes, and solid products. Additional probe attachments allow the use of the same analyzer for liquid transfer measurements.

Features:

Dual beam, pre aligned lamp assembly
No moving parts
Fast scanning speed -16000 wavelengths per second
The diameter of the sampling area at a sampling distance of 40mm is 6mm (the sampling amount can be adjusted)

Technical Specifications

Spectral range options: 6:00-1100 nm, 850-1700 nm, 900-1800 nm, 1100-2300 nm
Measurement mode: diffuse reflection, liquid transfer with probe attachment
Spectral resolution: 2-10 nm
Wavelength accuracy: ± 0.5 nm
Wavelength repeatability: ± 0.01 nm
S/N within 70% range: reflectivity and transmittance<30 μ abs, integration time<5 seconds
Wavelength access time:<66 μ sec
Luminous range: 3.5 AU

Signal digitization 1: 6-digit A/D (1 part in 65536)
Sampling speed: 16000 wavelengths per second
Sampling area: Ø 6mm Sampling distance 40mm
Power requirements: 12VDC, (24VDC special order), 90Watts, 110VAC 60Hz, 220VAC 50Hz
Output PC interface through Ethernet connection
Software package based on Windows data collection and analysis software