2B Model 205 Ozone Detector Ozone Analyzer from the United States


Product Introduction
·2B Model 205 Ozone Detector Ozone Analyzer from the United StatesyesThe 205 type dual ultraviolet ozone analyzer from 2B company in the United States adopts the 254nm ultraviolet dual beam light source technology and design of * * * new technology
·It can simultaneously measure the ozone value of the tested sample and the air value of the ozone free essence
·Especially suitable for measuring atmospheric ozone in space and on balloons
·The instrument range ranges from 1ppbv to 100ppmv. This instrument has high precision (1 ppb), small size, light weight (5 pounds, 2.3 kg), and low power consumption
Product Features
·Adopting dual beam 254nm ultraviolet light absorption method for high precision and resolution, with a range from 1ppbv to 100ppmv and a response time of 2 seconds
·Compared to similar products, it has lower battery consumption and can operate with an external power supply of 12V DC (5.0W)
·DewLineTM technology reduces interference from water vapor, with built-in data recording function and real-time clock display
·There are 3 sets of simulated signal outputs using RS-232 to output time, date, ozone concentration, instrument temperature, air pressure, and 3 additional sets of supplementary wheel input signals
·Exceptional external temperature, humidity, and air pressure analog output 0-2.5V, optional GPS function and CF memory card
Product Application
·Atmospheric measurement of balloons and aviation
·Monitoring in industrial production
·Establish a city monitoring network
·Personal and work environment monitoring
Standard Configuration
·Ozone Monitoring Model 205
·AC power adapter (specific plug for 100-240 VAC to 12V DC)
·Battery wire connector
·Zeroing ink cartridge
·Instrument Birth Certificate
·Operating Manual CD
·Calibration data and NIST traceable calibration certificates
·One year warranty
Product Parameters:
| Detection Principle | 254nm UV absorption technology with dual beams |
| Testing Scope | 1.0 ppb to 250 ppm |
| precision | 0.1ppb |
| Error (1 σ; rms noise) | ***Greater than 1.5ppb or 2% of the reading |
| threshold value | 3.0 ppb |
| NIST traceable calibration | have |
| Measurement interval | 2 seconds (average data options: 2 seconds, 10 seconds, 1 minute, 5 minutes, 1 hour) |
| traffic | 1.8 liters per minute |
| Traffic demand | >1.2 liters per minute |
| Baseline drift | <2 ppb/ 天 <5 ppb/year |
| Sensitivity drift | <1%/day <3%/year |
| Measure time and frequency | 10 seconds, 0.1 Hz |
| response time | 20 seconds |
| average time | 10 seconds, 1 minute, 5 minutes, 1 hour |
| data storage | 14,336 lines (10 s avg. = 1.4 days; 5 min avg = 1.4 mo.) |
| Ozone unit | At ppb, ppm, pphm, Microgram meters -3 meters -3 milligrams |
| weight | 4.7 pounds (2.1 kilograms); 1.6 pounds (0.7 kilograms) |
| size | 9cm x 21cm x 29cm |
| power supply | 12VDC, 5.0W (in low-power mode) |