1. The instrument installation adopts flange connection, threaded connection and clamping type;
2. The direction of liquid flow during installation should be consistent with the direction indicated by the arrow on the sensor housing, and the upstream straight pipe section should be ≥ 20dn, and the downstream straight pipe section should be ≥ 5dn (dn is the inner diameter of the pipeline).
3. Sensors should be kept away from external magnetic fields. If unavoidable, necessary measures should be taken;
4. In order to avoid affecting the normal transportation of liquids during maintenance, bypass pipelines should be installed outside the straight pipe sections at both ends of the sensor (as shown in Figure 4);
5. When installing sensors outdoors, please ensure that the amplifier plug is waterproofed;
6. The wiring of sensors and display instruments should be selected according to the power supply of the amplifier. Please refer to the relevant "User Manual" for details.
Liquid turbine flowmeter is a precision flow measurement instrument that can be used to measure the flow rate and total amount of liquid when paired with corresponding flow integrator instruments. Liquid turbine flow meters are widely used in measurement and control systems in fields such as petroleum, chemical, metallurgical, and scientific research. Liquid turbine flow meters equipped with sanitary fittings can be applied in the pharmaceutical industry.