(1) Display normal, settings normal, unable to heat up.
Solution to the problem: Connect the power supply, turn on the power switch, adjust the set temperature to be higher than the actual measured temperature, check if the temperature controller has an output indication. If there is, measure if there is voltage input to the heating tube. If there is, the heating tube is broken and can be replaced. If there is no voltage input to the heating tube, it is mostly due to relay failure. If the temperature controller does not have voltage output indication, it is recommended to replace the temperature controller.
(2) The temperature controller displays LLL or 999 (parameters may vary depending on the manufacturer)
The fault is a temperature sensor open circuit or short circuit, replace the sensor.
(3) Uncontrolled temperature constant temperature water bath pots are widely used for drying, concentration, distillation, impregnation of chemical reagents, impregnation of drugs and biological agents, as well as for constant temperature heating and other temperature tests in water baths. They are the * tool for education and scientific research in biology, genetics, viruses, aquaculture, environmental protection, medicine, hygiene, biochemical laboratories, and analysis laboratories Its main features: A. The studio water tank is made of stainless steel, which has excellent corrosion resistance B, Accurate temperature control, digital display, automatic temperature control C, Easy to operate and safe to use.