Cyclotron oscillatorIt is a biochemical instrument that can both cultivate and prepare biological samples, and is suitable for small-scale production. It is a laboratory equipment for scientific research, education, and production departments such as plants, biology, microorganisms, biological products, genetics, viruses, medicine, and environmental protection.
Usage:
1. First, place the container (bottle) that needs to be shaken into the oscillation plate spring. The solution in the container cannot exceed one-third of the container, otherwise it must be plugged
2. Check the timing knob and turn it to the ON (normally open) position or the desired time position. 3. Turn the speed control knob counterclockwise to the low-speed position, turn on the power switch, the power indicator light will turn on, and then turn the speed control knob clockwise to the desired speed. Precautions: 1. This rotary oscillator uses a three eye safety socket and must be properly grounded when in use. 2. When working, the machine should be placed on a flat and sturdy table to prevent vibration
3. The clamp spring is only used for fixing small and medium-sized bottles, and for extra large containers, several springs should be removed or reinforced in other ways
4. After using this machine, the power should be turned off, placed in a dry and ventilated place, and kept clean.
Maintenance points:
1. Proper use and maintenance of the rotary oscillator can prolong its service life by keeping it in good working condition.
During continuous operation, the rotary oscillator should be checked every three months to ensure that no water droplets, dirt, or other debris have fallen onto the motor and control components; Check the water quality and cleanliness inside the water tank; Check fuses, control components, and fastening screws
3. The bearings of the transmission part have been filled with an appropriate amount of lubricating grease (No. 1- calcium sodium based) before leaving the factory. During continuous operation of the instrument, lubricating grease should be added every six months, with a filling amount of about one-third of the bearing space.
4. After using the rotary oscillator, natural wear and tear is a normal phenomenon. After one year of use, the instrument found abnormal noise in the motor, bearing wear, loose or cracked belts, abnormal heating and constant temperature, electronic component failure, and other faults. Our company will continue to provide high-quality services and assist in handling them.