working principle
Online density/concentration meter uses a sound frequency signal source to excite a metal tuning fork and make it vibrate freely at the center frequency, which is related to the density of the liquid in contact. Therefore, by analyzing the frequency, the density of the liquid can be measured, and then temperature compensation can be carried out to eliminate the system's temperature drift; The concentration value at 25 ℃ can be calculated based on the relationship between the density and concentration of the corresponding liquid.
| Particle restriction | contain qi | Liquid viscosity | |
| particle diameter | suspended particles% | ||
| <10 microns | <=40% | recommend | 500cp |
| 10-50 microns | <=20% | Generally not used | 2000cp |
Application Industry
1. Petrochemical industry: diesel, gasoline, ethylene, etc.
2. Chemical industry: sulfuric acid, hydrochloric acid, nitric acid, ammonia water, methanol, ethanol, saline water, sodium hydroxide, frozen liquid, sodium carbonate, glycerol, hydrogen peroxide, etc.
3. Pharmaceutical industry: pharmaceuticals, biological fluids, alcohol extraction, acetone, alcohol recovery, etc.
4. Food and beverage industry: sugar water, fruit juice, brewing, cream, etc.
5. Battery and electrolyte industries: sulfuric acid, lithium hydroxide, etc.
6. Environmental protection industry: desulfurization (lime slurry, stone paste slurry), denitrification (ammonia water, urea), wastewater treatment (acid, alkali, salt recovery), etc.
Product dimensions

| precision | ±0.003g/cm3 | ±0.25% |
| scope of work | 0~2g/cm3 | 0~99% |
| repetitiveness | ±0.0005g/cm3 | ±0.1% |
| Process temperature impact (corrected) | ±0.0001g/cc | ±0.1kg/cm3 (Every℃) |
| work pressure | ≤ 2.5MPa (standard); ≤ 6.3MPa (customized) | |
| Test pressure | Under work pressure Test under 1.5 times pressure | |
| Liquid-contacting part | 316L、 Hastelloy, titanium alloy, zirconium, etc | |
| Fork tooth polishing | Standard polishing PTFE coating, electrolytic polishing | |
| shell | IP65, aluminum alloy | |
| Power requirements | 24 VDC,100mA | |
| output | Four wire system,4-20mA/RS485 | |
| electrical interface | M20*1.5 | |
| Process connection type | Flanges, clamps, threads | |
Fluid and material compatibility
name name molecular formula Concentration(%) 316L Hastelloy titanium alloy zirconium PTFE sour Hydrochloric acid(Excluding hydrofluoric acid HCI 0-40 × ○ × ☆ ☆ sulfuric acid H2SO4 0-50 ○ ☆ ○ ☆ ☆ H2SO4 50-75 × ○ × ○ ☆ H2SO4 75-98 ○ ☆ ○ ○ ☆ nitric acid HNO3 0-100 ○ ○ ○ ○ ☆ phosphoric acid H3PO4 0-98 × × × × ☆ alkali sodium hydroxide NaOH 050 ☆ ☆ ☆ ○ ☆ potassium hydroxide KOH 0-50 ☆ ☆ ☆ ○ ☆ calcium hydroxide Ca(OH)2 0-50 ☆ ☆ ☆ ○ ☆ salt sodium chloride NaCl 0-50 × ☆ ☆ ☆ ☆ ammonium chloride NH4Cl 0-50 × ☆ ☆ ☆ ☆ ammonium sulfate (NH4)2SO4 0-50 × ☆ ☆ ☆ ☆ other urea (NH2)2CO 0-100 ☆ ☆ ○ ☆ ☆ Sodium hypochlorite NaOCI 0-16 × ○ × ☆ ☆ hydrogen peroxide H2O2 0-90 ☆ ☆ ○ ☆ ☆ ☆ Recommended ○ Use under specific concentration and temperature restrictions Not used