Instrument Description: The touch sensitive COD ammonia nitrogen total phosphorus total nitrogen turbidity analyzer adopts a 7-inch full touch screen, flat icon design, guided operating system, and simple and easy to understand interface, allowing operators to use it quickly and accurately. The operation does not require manual intervention, and the system automatically identifies the detection channel for precise detection.
Instrument Introduction
The YC9100-4 touch sensitive COD ammonia nitrogen total phosphorus total nitrogen turbidity analyzer is designed and developed based on relevant national standards. It adopts a 7-inch full touch screen, flat icon design, guided operating system, and simple and easy to understand interface, allowing operators to use it quickly and accurately.
The detection system adopts an independent multi module optical (imported) design with excellent optical performance. The operation does not require manual intervention, and the system automatically identifies the detection channel for precise detection. The detection uses a 16mm digestion detection integrated optical colorimetric tube, simplifying the operation steps and avoiding errors caused by repeated liquid removal, greatly improving accuracy.
Technical Specifications
| Test Item | COD | ammonia nitrogen | Total Phosphorus | Total Nitrogen | turbidity | Suspended solids (SS) |
| Testing Standard | HJ/T399-2007 | HJ535-2009 | GB1183-89 | EPAColor changing acid method | Determination of turbidity in water quality | Improvement of National Standard Weight Method |
| measurement range | 0-15000mg/L | 0-50mg/L | 0-20mg/L | 0-100 mg/L | 0-1000NTU | 0-500mg/L |
| detection limit | 5mg/L | 0.01mg/L | 0.02mg/L | 0.5mg/L | 0.5mg/L | 5mg/L |
| Dissolve the environment | 165℃,20min | No need to dissolve | 125℃,30min | 125℃,30min | No need to dissolve | No need to dissolve |
| Instrument model | YC9100-4 |
| Measurement error | ≤± 5% |
| repetitiveness | ≤±3% |
| Optical stability | ≤±0.001A/20Minutes(100000Hour without decay) |
| Colorimetric method | Colorimetric tube (digestion colorimetric integrated tube) |
| data processing | 3Ten thousand records96Curve (supports custom extension projects) |
| transmission method | USBLocal Area Network |
| printer | Built in thermal printer |
| operating interface | Touchscreen Chinese/English (optional) |
| display screen | 7Inch color touch screen |
| power supply method | AC(220V±10%),50Hz |
| power | 5W |
| working environment | 5-45℃ ≤85%No condensation |
| host size | 348mm*250mm*145mm |
| Instrument weight | host<2.6kg; Dissolver<7kg |
Technical parameters of digestion instrument
| Dissolve sample size | 25branch | time display | Real time date and time display |
| operating interface | Chinese display | Operation method | Button type |
| rated voltage | AC 220V±10%/50Hz | Power requirements | grounding |
| Dissolve pore size/depth | 16mm/65mm | Dissolving capacity | >12mL |
| display screen | TFTHigh definition color screen | timing error | 10min±0.1s |
| Pre set digestion mode | CODTotal phosphorus, total nitrogen | Dissolution method | Fully enclosed type |
| safety protection | Double protection against overheating | temperature control | intelligencePIDtemperature control |
| Custom Mode | 5The temperature and time can be set arbitrarily | Project switching | Quick switch |
| Reminder Method | buzzer | Alarm Alert1 | Reaching temperature and maintaining constant temperature |
| Custom Time | 5-999minadjustable | Alarm Alert2 | Dissolve completed and stop heating |
| Custom Temperature | 90~200℃ adjustable | Dissolve protection | thermal insulation board/protective shield |
| Temperature control accuracy | <±0.5℃ | Temperature control sensor | Imported PlatinumPT100 |
| heating rate | Set to165℃,12Constant within minutes | Instrument weight | 6kg |
| working environment | -10-40℃, relative humidity<85% | Overall dimensions | 230*340*130(mm) |
Example of operating interface
Instrument Features
1. Full touch design: 7-inch touch screen, supporting Chinese/English display (optional), guided operating system, simple and time-saving.
2. Wide detection range: dozens of detection parameters can be expanded, and customized according to customer needs.
3. Detection optics: Adopting independent multi module optical (imported) design, excellent optical performance, stable and reliable.
4. Operating System: Guided operating system with a simple and easy to understand interface, guiding users to complete operations quickly and accurately.
5. Operating program: a. Built in system program and user program, which can add frequently tested items to the user program for easy searching.
b. It has power-off protection function and one click factory reset function to prevent data loss.
c. The built-in standard work program can be used directly, and users can also expand the detection items themselves.
d. Built in help interface, which can query user detection process problems and issues.
e. It has a one click factory reset function, which can quickly recover data loss in case of unexpected situations.
6. Intelligent detection: Standard measurement and continuous measurement can be switched freely, reducing repeated measurement steps and improving detection efficiency.
7. Operation safety: Adopting a digestion colorimetric integrated tube, reducing the steps of transferring reagents, making the measurement simple, fast, and safe.
8. Quick printing: Equipped with a built-in thermal printer, it can print both current and historical measurement data.
9. Data storage: Standard working curve and 96 user-defined items, capable of storing 30000 pieces of detection data.
10. Online function: Supports data transfer between network port, LAN, USB and PC or printer.
11. Professional reagents: specialized reagents, simplified operating procedures, reduced operational errors, less reagent usage, and lower consumable prices.
National Unified Service Hotline: