Measurement principle:
Two electric heating regulators connected for different measurements are suspended in a constant temperature measuring cell filled with saturated vapor of solvent. The initial surface of the measuring probe is covered with solvent droplets, so there is no temperature difference between them. Replacing a solvent droplet on the surface of a probe with a solution droplet can cause solvent vapor condensation due to the lower solvent vapor pressure on the solution surface. The condensation release causes the temperature of the solution droplets to increase, while also leading to an increase in vapor pressure. When the saturated vapor on the surface of the measuring cell and solution droplets reaches pressure equilibrium, the solution droplets reach a relatively stable temperature, which ultimately leads to a difference in temperature between the two probes, and the temperature difference is directly converted into a voltage signal. The measured value is proportional to the osmotic concentration of the solution.
Due to the nonlinear behavior of osmotic pressure and concentration in polymer solutions, it is necessary to measure solutions with at least three concentration gradients. Calculate measurement results using statistical methods such as linear regression.
◆OSMOMAT 070Application of vapor pressure osmometer:
Suitable for determining the average molecular weight of polymers dissolved in different organic solvents and the osmotic pressure of aqueous solutions. The measurement range is40-50000g/molOrganic solubility
Product Model
parameter/model |
OSMOMAT 070 |
OSMOMAT 090 |
measurement range |
Zui low measurement concentration:3.3 x 10-5 Mol/kg(Toluene)1.7 x 10-4 Mol/kg(Water) |
Zui small detection pressure0.1 mmSolvent column |
Molecular weight range |
Toluene:50 - 50.000Dalton |
5000-2.000.000Dalton, (depending on the chosen semi permeable membrane) |
sample volume |
about1 ml/Each solution should have at least3A solution (a solution with a molar concentration gradient) |
Measurement pool volume: approximately5ml |
Measuring principle |
Measure the vapor pressure of the sample:Less than the solvent used0.1% |
|
calibration material |
Diphenylethanone/Diphenylethyl alcohol/Naphthalene, with a molar concentration of0.1—1.5Osmol/kg(water)or1 x 10-3—3 x 10-2(organic solution) |
|
Probe for measuring two temperature ranges |
Paired insertion (water and organic solution) |
|
Membrane diameter |
|
40mm |
solvent |
|
Various organic solvents, water, and electrolyte solutions containing water |
analog output |
1mV/Number, the largest10V |
|
temperature adjustment |
1℃scale |
1℃scale |
temperature range |
High room temperature5oC-130oC |
High room temperature5oC-130oC |
size(HXD) |
260x320x230mm |
260x320x230mm |
weight |
approximately9.3kg |
approximately10kg |
power supply |
180VA |
180VA |