E-mail
sales@cspmk.com
- Phone
-
Address
6th Floor, Jinyanlong Building, Xisanqi, Haidian District, Beijing
Beijing Speke Technology Co., Ltd
sales@cspmk.com
6th Floor, Jinyanlong Building, Xisanqi, Haidian District, Beijing
simple be situated between
SPMK518 intelligent process calibrator(Multi functional temperature and pressure process calibrator)use3.5cunTFTColor screen, Chinese display, mobile menu operation mode, built-inHARTHandheld controller, with powerful task management functions, meets the requirements of calibrating temperature, pressure and other process instruments in industrial sites and laboratories. It has a beautiful appearance, sturdy structure, compact and economical practicality, making it an ideal tool for calibrating, maintaining and repairing thermal instruments.
main features
◆ It can measure voltage, current, resistance, frequency, pulse, switch (on/off), and can also use thermal resistors or thermocouples to measure temperature without interfering with each other;
◆ Can simulate13Thermocouples and11Measure and output thermal resistance, and provide cold end compensation;
◆ Provide isolation between measurement and output24VDCPower Supply;
◆ Support data storage function, support step or programming output methods;
◆ Built inHARTHandheld device, capable of achievingHARTManagement of basic information of type instruments, migration of measuring ranges, calibration and maintenance of functional equipment such as zero and full calibration;
◆ ConnectableSDPModular intelligent digital pressure calibrator;
◆ Provide commonly used tools such as thermal engineering encyclopedia, analog transmitter, user built thermal resistance library, temperature and pressure unit conversion, etc;
◆ Automatic cold end compensation technology: embedded cold end insulation module, quickly tracks temperature changes;
◆3.5Inch high resolutionTFTDisplay screen, Chinese language display, good human-computer interaction interface;
◆RS232The interface and open communication protocol facilitate remote networking for users and support online firmware upgrades.
Technical indicators for signal measurement(Environmental temperature)20℃±5℃,One year accuracy
Signal type |
range |
accuracy |
resolution |
|
electric current |
(-30~30)mA |
±(0.01%RD+0.005%F.S) |
0.1uA |
|
voltage |
(-30~30)V |
±(0.01%RD+0.005%F.S) |
0.1mV |
|
Millivolt voltage |
(-75~75)mV |
±(0.01%RD+0.005%F.S) |
0.1uV |
|
resistance |
3Line system |
(0~400)Ω |
±(0.02%RD+0.005%F.S) |
1mΩ |
4Line system |
(0~400)Ω |
±(0.01%RD+0.005%F.S) |
1mΩ |
|
frequency |
(1~50000)Hz |
±(0.005%RD+0.002%F.S) |
1Hz |
|
pulse |
0~999999 |
±1A pulse |
1 |
|
Switch (on/off) |
If the switch is equipped with inspection voltage, the voltage range is(3~24)V |
|||
thermocouple |
K、E、J、T、N、R、S、B、C、D、G、L、U |
|||
thermal resistance |
Pt1000(385)、Pt500(385)、Pt100(385)、Pt100(3916)、Pt100(3926)、Pt10(385)、Cu10(427)、Cu50(428)、Cu100(428)、Ni120(672)、Ni100(618)And custom types |
|||
pressure |
Refer to this websitePressure Range Selection Table |
|||
Technical specifications for signal output(Environmental temperature)20℃±5℃, one-year accuracy)
Signal type |
range |
accuracy |
resolution |
electric current |
0~22)mA |
±(0.02%RD+0.005%FS) |
0.1uA |
voltage |
0~12)V |
±(0.02%RD+0.005%FS) |
0.1mV |
Millivolt voltage |
(-10~100)mV |
±(0.02%RD+0.005%FS) |
0.1uV |
frequency |
0~50000.0)Hz |
±(0.005%RD+0.002%FS) |
0.1Hz |
pulse |
0~999999 |
±1A pulse |
1 |
resistance |
0~400)Ω |
±(0.02%RD+0.005%FS) |
1mΩ |
(400~4000)Ω |
±(0.03%RD+0.01%FS) |
10mΩ |
|
24VDC |
(24 ± 0.5) V maximum load capacity 50mA |
||
thermocouple |
K、E、J、T、N、R、S、B、C、D、G、L、U |
||
thermal resistance |
Pt1000(385)、Pt500(385)、Pt100(385)、Pt100(3916)、Pt100(3926)、Pt10(385)、Cu10(427)、Cu50(428)、Cu100(428)、Ni120(672)、Ni100(618)And custom types |
||
Other indicators
1. Operating environment: ambient temperature (-10~50) ℃; Relative humidity<90% non condensing; Atmospheric pressure (86~106) KPa. a.
2. Storage environment: ambient temperature (-20~60) ℃; Storage humidity<90% non condensing.
3. Power supply: AC/DC power supply.
Note: In order to achieve better measurement and output results for the calibrator, it is recommended to use battery power.
packing list
name |
model |
unit |
quantity |
Multi functional process calibrator |
SPMk518 |
tower |
1 |
operation instructions |
《SPMK518User Manual |
stem or root of plants |
1 |
warranty card |
share |
1 |
|
certificate of conformity |
share |
1 |
|
The power adapter |
individual |
1 |
|
Pen line |
0.5rice red |
strip |
1 |
Pen line |
0.5rice black |
strip |
1 |
Pen line |
1rice red |
strip |
3 |
Pen line |
1rice black |
strip |
3 |
Alligator Clip |
red |
individual |
3 |
Alligator Clip |
black |
individual |
3 |
Compensation wire |
Ktype |
strip |
1 |
Compensation wire |
Etype |
strip |
1 |
The signal line |
Seven Core |
strip |
1 |
Compensation wire |
Utype |
cover |
1 |
packing list |
share |
1 |