purpose |
|
|
ZY-3051GP remote pressure transmitter is a device installed on a pipeline or container to sense the measured pressure. The pressure is transmitted to the transmitter body through a capillary filled with silicone oil (or other liquid), and then converted to 4-20mA by the delta chamber and amplification circuit board inside the transmitter body DC signal output (see schematic diagram). The TS-3051GP remote pressure transmitter can be combined with an intelligent amplification board to form an intelligent remote pressure transmitter. When paired with a handheld device that complies with the HART protocol, it can communicate with each other for setting and monitoring. In order to meet different installation requirements, this series of transmitters has multiple forms of remote transmission devices for users to choose from, and the main structure of the transmitter is the same as that of a pressure transmitter. |
|
|
|
|
Remote transmitters are mainly used for measurement in the following situations |
|
|
Viscous medium at high temperature Easily crystallizable medium Settling media with solid particles or suspended solids Strong corrosive or highly toxic media Can eliminate the occurrence of pressure pipe leakage and pollution of the surrounding environment; It can avoid the tedious work of frequently replenishing isolation fluid due to the instability of measurement signals when using isolation fluid. Continuous and precise measurement of interface and density The remote transmission device can avoid the mixing of different instantaneous media, thus enabling the measurement results to truly reflect the actual situation of process changes. Places with high hygiene and cleanliness requirements In the production of food, beverage, and pharmaceutical industries, not only is it required that the contact area of the transmitter with the medium meet hygiene standards, but it should also be easy to rinse to prevent cross contamination of different batches of media. |
|
|
Product Features |
|
|
Simulate "online operation" assessment for each transmitter to ensure stable and reliable operation under extreme environmental temperature, medium temperature, and working pressure (including positive pressure or vacuum). Adopting a "dynamic profile" remote transmission membrane structure, the membrane will not be damaged after overload; The accuracy will not change after long-term work. The use of a full fusion welding and rigid sealing structure filling system has eliminated the occurrence of oil leakage. A novel capillary coupling component is designed to provide reliable protection for the capillary, making it less prone to breakage during on-site use. A variety of remote transmission devices in terms of variety, form, and size, as well as a rich variety of filling fluids, can meet the measurement needs of various occasions. We can provide various special needs products, including high-temperature, high vacuum, fast response, and unequal long capillary (length ≤ 3 meters) remote transmission pressure transmitters. |
|
Technical parameters and performance |
|
|
Output signal: 4-20mA DC two-wire system (analog) The two-wire 4-20mA DC signal is overlaid with a digital signal, and the user selects linear or square root output. (Intelligent) Power supply: 12-45V DC (see load characteristic diagram for details) Load characteristics: Power supply impact:<0.005% output range/V Load impact: When the power supply is stable, there is no load impact. Measurement accuracy: ± 0.1% and ± 0.2% of the calibration range (The standard precision is ± 0.2%. If you choose other precision, please indicate it when ordering and selecting.) Range ratio: 10:1 or 100:1 Damping: It is usually adjustable between 0.1 and 16 seconds, and the time constant increases when filled with inert fluid or with a remote transmission device. Start time:<2 seconds, no preheating required |
|
|
Working environment: ambient temperature -29~93 ℃ (analog amplifier) -29~75 ℃ (digital/smart amplifier) -29-65 ℃ (with display header) Environmental humidity: 0-95% Protection feature: IP65 protection capability Explosion proof type: Explosion proof Exd II BT4-6 Intrinsically safe Exia II CT5 Static pressure influence: Zero position error: ± 0.5% maximum range limit, calibrated by zeroing for 32MPa under pipeline pressure. Electromagnetic radiation impact: 0.05% maximum range value, accepted radiation frequency of 27-500MHz, test field strength of 3V/m. Indicator (%): LCD digital display accuracy ± 0.2% Vibration effect: ± 0.05%/g when vibrating at 200Hz in any direction. Installation position: The diaphragm is not installed vertically, which may result in a zero point error of less than 0.24KPa, but it can be corrected by adjusting the zero point. Weight: 3.9Kg (excluding accessories)
pourThe working pressure of the remote pressure transmitter also depends on the specifications of the selected flange. ② The working temperature of the remote transmission device is determined by the type of filling fluid selected. ③ A remote pressure transmitter refers to the output change caused by the temperature effect on one side of the remote transmission device. ④ The effects of static pressure and temperature were measured at the maximum range. ⑤ The length of the capillary tube on one side of the GP type remote pressure transmitter is generally 1.5 meters, and the longest should not exceed 3 meters (agreed upon by the protocol in special circumstances).
|
|
|
Flat flange (RFW) remote pressure transmitter |
|
|
|
|
Convex flange (EFW) remote pressure transmitter |
|
|
|
|
Flat type (PFW) remote pressure transmitter |
|
|
|
|
|
Dimensional Table of Connection Flanges for Remote Transmission Devices (Applicable to TS-RFW, TS-EFW, and TS-PFW Types) |
|
|
Company: mm
|
|
|
|
Flange specifications
|
Flange size
|
Pad ring
|
Installation size
|
Insert cylinder
|
|
code
|
orifice
|
pressure grade
|
outside diameter D
|
Convex diameter d
|
Convex height f
|
thickness
|
diameter F
|
aperture φ
|
Distribution circle K
|
Number of holes n
|
diameter h
|
diameter B
|
Protruding length L
|
|
A
|
E
|
|
0A1
|
1"
|
150Lb
|
108
|
___
|
5.5
|
20
|
24
|
66.5
|
27
|
79.4
|
4
|
15
|
___
|
___
|
|
0A3
|
300Lb
|
124
|
23
|
88.9
|
19
|
|
0A6
|
600Lb
|
|
1A1
|
11/2"
|
150Lb
|
127
|
79
|
42
|
98.4
|
15
|
|
1A3
|
300Lb
|
156
|
26
|
114.3
|
23
|
|
1A6
|
600Lb
|
28
|
|
2A1
|
2"
|
150Lb
|
152
|
92
|
23
|
92
|
52
|
120.6
|
19
|
Temporary supply
60
|
50 (2") 100(4") 150(6")
|
|
2A3
|
300Lb
|
165
|
26
|
127
|
8
|
|
2A6
|
600Lb
|
31
|
|
3A1
|
3"
|
150Lb
|
191
|
127
|
27.5
|
127
|
79
|
152.4
|
4
|
19
|
77
|
|
3A3
|
300Lb
|
210
|
32.5
|
168.3
|
8
|
23
|
|
3A6
|
600Lb
|
37.5
|
|
4A1
|
4"
|
150Lb
|
229
|
157
|
28
|
157
|
103
|
190.5
|
19
|
95
|
|
4A3
|
300Lb
|
254
|
33.5
|
200
|
23
|
|
4A6
|
600Lb
|
273
|
43.5
|
216
|
25
|
|
|
Threaded Installation (RTW) Pressure Transmitter |
|
|
|
|
Sanitary (SSW) remote pressure transmitter |
UCW remote pressure transmitter with threaded connection |
|
|
|
|
|
Form size (mm) |
Connection thread diameter D
|
|
M33×2 (G1")
|
M48×3 (G1/2")
|
M60×3 (G2")
|
|
Hexagonal opposite side S
|
36
|
50
|
63
|
|
Thread length A
|
20
|
30
|
30
|
|
Total length C
|
40
|
60
|
70
|
|
Probe diameter d
|
30
|
44
|
55
|
|
Diaphragm diameter B
|
24
|
35
|
44
|
|
Minimum allowable range (KPa)
|
1000
|
300
|
100
|
|
|
|
Note: 1. Can provide connection threads of specified size. 2. The protrusion length L is available in three types: 0.5 "(12.5mm), 1" (25mm), and 2 "(50mm). |
|
Selection of remote pressure transmitter |
|
|
| Remote transmission transmitter is a variant product that adds a remote transmission device to the transmitter body, so its selection work can be carried out in two steps: first, determine the type (DP), measurement range, and structural materials of the transmitter body, and then determine the type and size of the remote transmission device according to the installation requirements on site. Due to the close relationship between the various parameters that make up the filling system (such as the working temperature, viscosity, and capillary length of the filling fluid) and the performance indicators of the transmitter, they are included in the selection category of the transmitter body and are not separately listed for selection. |
|
|
TS-3051GP remote pressure transmitter |
|
|
| Code measurement range (KPa) |
|
4
5
6
7
8
9
0
|
0 - 6.2 0 - 31.1 0 - 117 0 - 345 0 - 1170 0 - 3450 0 - 6895
|
~ 37.4 ~ 186.8 ~ 689.5 ~ 2068 ~ 6895 ~ 20680 ~ 41370 |
pour1. Remote transmission devices PFW, EFW, RTW, SSW, and RFW can choose range codes 5, 6, 7, and 8. 2. The remote transmission device UCW can choose range codes 6, 7, 8, 9, and 0. |
|
|
|
| Code output format |
|
SF S
|
4~20mA. DC digital amplifier with 4-digit digital display 4~20mA. DC intelligent amplifier and uses HART protocol communication |
|
|
|
| Code capacity and joint material, isolation membrane material |
12 carbon steel 316 steel 22 316 steel 316 steel |
pourGenerally, code 12 is selected, and code 22 is chosen in humid and corrosive environments. |
|
|
|
| Code: Type of filling fluid |
Low viscosity silicone oil -60-150 ℃ ρ=0.93 S ordinary silicone oil -45~205 ℃ ρ=0.96 High temperature silicone oil -10~300 ℃ ρ=1.07 N vegetable oil -15~200 ℃ ρ=0.92 Distilled water+5~70 ℃ ρ=1.0 |
pourThe codes N and W are used in situations with strict hygiene requirements to avoid contamination of process media caused by leakage of filling fluid. |
|
|
|
| Code capillary length (m) |
05 1.5(5ft) 10 3.0(10ft) |
pourWhen there are no special requirements, please fill in the code 05, which means the capillary length is 1.5 meters. |
|
|
|
| Code selection parts |
M3 digital indicator with 4-digit LCD digital display B1 pipe bending bracket (installation pipe diameter 60) B2 board bending bracket B3 pipe installation flat bracket (installation pipe diameter 60) D explosion-proof type, explosion-proof mark Exd Ⅱ BT4-6 I intrinsic safety type, explosion-proof mark Exia Ⅱ CT5 |
|
|
|
|
|
|
| ZY-3051GP 7E 22 S 05 M1 B3 0~2MPa Selection Example |
|
|
XY-RFW type flat flange remote transmission device |
|
|
| Code washing spare hole |
11 None 21 has |
pourWhen the flange diameter is less than 2 ", a gasket ring is required. When the flange diameter is greater than or equal to 2", a gasket ring is only required if there is a flushing hole. |
|
|
|
| Code: Remote Transmission Diaphragm Material |
316L steel B Hastelloy C Tantalum C N Nickel 20 V Monel alloy H Titanium |
|
|
|
| Code installation flange diameter |
| 1 1" |
|
|
|
| Code Installation Flange Standard |
| A ANSI or HG20615 |
|
|
|
| Code installation flange pressure rating |
1 150Lb or PN10/16bar 3 300Lb or PN25/40bar 6 600Lb or PN64bar |
|
|
|
| Code: gasket ring material, loose flange material, diaphragm seat material |
A 316 steel, 304 steel, 316 steel B Hastelloy C 304 Steel 316 Steel E 316 steel carbon steel 316 steel |
pourStructural material code B, equipped with remote transmission diaphragms other than 316L steel. |
|
|
|
|
|
|
| TS-RFW 21 C 3 A 3 B Selection Example |
|
|
TS-EFW type convex flange remote transmission device |
|
|
| Code installation flange diameter |
2 2" 3 3" 4 4" |
pourWhen selecting range code 3, a diameter of 4 "must be used. |
|
|
|
| Code Installation Flange Standard |
| A ANSI or HG20615 |
|
|
|
| Code installation flange pressure rating |
1 150Lb or PN10/16bar 3 300Lb or PN25/40bar 6 600Lb or PN64bar |
|
|
|
| Code: Remote Transmission Diaphragm Material: Insertion Cylinder Material: Loose Sleeve Flange Material |
316L steel 316 carbon steel B Hastelloy C 316L Steel 304 Steel C Tantalum 316L Steel 304 Steel |
pourIn damp and corrosive environments, use 304 steel flanges. |
|
|
|
| Code diaphragm protrusion length (mm) |
2 50(2") 4 100(4") 6 150(6") |
pour: Available in non series lengths. |
|
|
|
|
| ZY-EFW 3 A 1 A 8 Selection Example |
|
|
XY-PFW Flat Remote Transmission Device |
|
|
| Code washing spare hole |
11 None 12 has |
pourThe flushing hole is located on the gasket ring. |
|
|
|
| Code: Remote Transmission Diaphragm Material |
316L steel B Hastelloy C Tantalum C N Nickel 200 Y Monel alloy H Titanium |
|
|
|
| Code installation flange diameter |
2 2" 3 3" 4 4" |
pourWhen selecting range code 3, a diameter of 4 "must be used. |
|
|
|
| Code Installation Flange Standard |
| ANSI and HG20615 |
|
|
|
|
| Code installation flange pressure rating |
1 150Lb or PN10/16bar 3 300Lb or PN25/40bar 6 600Lb or PN64bar |
pour: Refers to the pressure rating for installing accessory flange covers. |
|
|
|
| Code: gasket ring, flange cover, diaphragm seat |
|
A B E
|
316 steel Hastelloy C 316 steel
|
304 steel 304 steel carbon steel
|
316 steel 316 steel 316 steel
|
pourStructural material code B, equipped with remote transmission diaphragms other than 316L steel. |
|
|
|
|
|
|
| Example of ZY-PFW 21 A 4 A 3 E Selection |
|
|
ZY-RTW type threaded installation remote transmission device |
|
|
| Code flushing hole |
11 None 12 has |
|
|
|
| Code: Remote Transmission Diaphragm Material |
316L steel B Hastelloy C Tantalum C |
|
|
|
|
| Code: Installation flange material, sealing gasket material, membrane sheet seat material |
11
31 |
carbon steel 304 steel |
PTFE PTFE |
316 steel 316 steel
|
pourStructural material code 31 is equipped with remote transmission diaphragms other than 316L steel.
|
|
|
|
| Code: Liquid receiving cover |
|
A B E
|
316 steel Hastelloy C carbon steel
|
pourMaterial code B for the liquid receiving cover, equipped with a remote transmission diaphragm other than 316L steel.
|
|
|
|
|
|
|
|
|
|
| ZY-RTW 11 A 31 A 15 Selection Example |
|
|
TS-SSW type sanitary remote transmission device |
|
|
| Code insertion tube |
31 diameter 3 "(Φ 70), pressure rating 150LB (2MPa) 33 diameter 3 "(Φ 70), pressure rating 300LB (5MPa) |
|
|
|
| Code: Remote Transmission Diaphragm Material: Insertion Tube Material: Clamp Material |
|
A B
|
316L steel Hastelloy C
|
316 steel 316 steel
|
|
|
|
|
| Code protrusion length (mm) |
|
2
4
6
|
50(2") 100(4") 150(6") |
|
|
|
| Code installation attachment |
|
N Y1 Y2
|
without With clamps Equipped with clamps and installation cans |
pourCode N and Y1 are used for spare parts supply, while code Y2 is used for new equipment installation. |
|
|
|
|
|
|
| Example of TS-SSW 31 A 4 Y2 Selection |
|
|
XY-UCW type threaded connection remote transmission device |
|
|
| Code O-ring material |
|
11 |
PTFE |
|
|
|
| Code: Remote Transmission Diaphragm Material: Probe Body Material: Hexagonal Bolt Material |
A B |
316L steel Hastelloy C |
316 steel Hastelloy C |
304 steel 316 steel |
|
|
|
| Code protrusion length (mm) |
05
10
20 |
12.5(0.5") 25(1") 50(2") |
pourThe maximum protruding length can reach 100mm. |
|
|
|
| Code connection thread |
|
A B C
|
M33 × 2 (G1 ") measurement range 0-1.0~42MPa M48 × 3 (G1 ") measurement range 0-0.3~21MPa M60 × 3 (G2 ") measurement range 0-0.1~7.0MPa |
pourOptional range code: 8.9.0
pourOptional range code: 7.8.9
pourOptional range code: 6.7.8 |
|
|
|
|
|
|
| Example of ZY-UCW 11 A 10 C Selection |
|
|
|
|