Power and Harmonic Analyzer

NegotiableUpdate on 03/18
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Numerous measurement functions, integrating power meter, harmonic meter, multimeter, and oscilloscope in one. Including true RMS voltage, current, power, power factor, and true power factor, up to the 51st harmonic, harmonic phase, and total harmonic distortion. : Zhang Yunhua;:
Product Details

F43B power and harmonic analyzerProduct IntroductionCircuit breaker action, transformer overheating, motor burnout, flashing lights, no matter what kind of power quality problem, Fluke's f43 power quality analyzer can help you find the problem. This sturdy and portable instrument can meet your needs for diagnosing power quality and general equipment faults. It is also an ideal tool for routine power maintenance and accident prevention. The F43 integrates the most useful power quality analyzer, multimeter, and oscilloscope into a sturdy, portable, and easy-to-use testing instrument. Simply select the measurement application from the main menu, and the rest will be automatically completed by the instrument. Numerous measurement functions, integrating power meter, harmonic meter, multimeter, and oscilloscope in one. Including true RMS voltage, current, power, power factor, and true power factor, up to the 51st harmonic, harmonic phase, and total harmonic distortion. High brightness display of voltage, current waveform or harmonic spectrum on a large screen. Monitor and record up to 40 voltage transient events. Up to 8 days of online recording and capturing intermittent changes. The multimeter function can measure resistance, diodes, capacitance, and on-off. Support temperature testing options, direct reading of temperature readings. Randomly provide software and RS232 interface cables for computer analysis and document processing. Menu operation, also learn and apply. The internal storage function can accommodate up to 10 screens.

F43B power and harmonic analyzerGeneral indicators

power supply
Line voltage adapter/charger
Interior Decoration Ni CD rechargeable battery
Usage time 4 hours
Charging time 4 hours
Environmental Requirements
Temperature 0 ℃ to 50 ℃
environment mil 28800e, type 3, class ⅲ, style b
Encapsulation IP51 (dustproof, waterproof drip)
Mechanical data
Appearance (h × w × d) 232×115×50mm
weight 1.1kg
safe
Compliant with IEC 61010-1 Class III 600V safety standard
Impact protection Input A and B, 6kV
Floating measurement 600V RMS, any terminal to ground
guarantee Host for three years, attachment for one year

Technical specifications: Accuracy is expressed as ± (percentage of reading+words), excluding probes. The character is the least important position in the display. For example, displaying 1.00v, each word represents 0.01v

input characteristics
range
accuracy
input impedance 1mω, 12pf
input voltage 600V true effective value, cat iii
Voltage/current/frequency
True RMS voltage (ac+dc) 5.000v, 50.00v, 500.0v, 1250v ± (1%+10 words)>1.4vrms
True effective current (ac+dc) 50.00a, 500.0a, 5.000ka, 50.00ka ± (1%+10 words)>3arms
main frequency 40.0 to 70.0Hz ± (0.5%+2 words)
Power display
Active power, watts apparent power, va reactive power, var 250w, 2.50kw, 25.0kw, 250kw,2.50mw ± (4%+4 words)
Power factor, pf displacement power factor, dff 0.00 to 0.250.25 to 0.90-0.90 to 1.00 not specified±0.04±0.03
F43BHarmonic display
voltage fundamental wave ± (3%+2 words)
Up to the 31st harmonic ± (5%+3 words)
Up to the 51st harmonic ± (15%+5 words)
power fundamental wave ± (5%+10 words)
Up to the 31st harmonic ± (10%+10 words)
Up to the 51st harmonic ± (30%+5 words)
frequency fundamental wave ±0.25hz
phase Fundamental to 51st harmonic ± 3 ° to ± 15 °
K-factor 1.0 to 30.0 ±10%
voltage recording
Record time 4 minutes to 16 days optional
True RMS voltage 5.000v, 50.00v 500.0v, 1250v ± (2%+10 words)
True effective current 50.00a, 500.0a, 5.000ka, 50.00ka ± (2%+10 words)
other records
Record time 4 minutes to 16 days
parameter Select 2 from the following parameters
v/a/hz Voltage, current, frequency
power watts, va, var, dpf, frequency
harmonic Total harmonic distortion (thd), harmonic voltage (current, power), frequency, phase
resistance Resistance, diode, on-off, capacitor
temperature temperature
oscilloscope AC/DC voltage, AC/DC current, frequency, pulse width, phase, duty cycle, maximum and minimum peak values, peak difference, peak factor
Transient event capture
Zui small pulse width 40ns
Voltage threshold setting Normal values of 20%, 50%, 100%, and 200% above and below
The cursor position is the smallest and the voltage is the largest 10v, 25v, 50v, 125v, 250v, 500v, 1250v ± 5% full scale
inrush current
current range 1a, 5a, 10a, 50a, 100a, 500a, 1000a
impact time 1s, 5s, 10s, 50s, 100s, 5min
Cursor reading 12.5a, 25a, 50a, 125a, 250a 500a, 1250a, 2500a, 5ka ± 5% full scale
Cursor time 4 to 235 pixels * ± (0.2%+2 pixels)
Oscilloscope display
Time range 20ns/grid to 60s/grid
The maximum sampling rate 25ms/s
bandwidth Voltage channel [1] Current channel [2] 20mHz, TL test line at 1mHz 15kHz, 80i-500s probe at 10kHz
coupling ac, dc
Vertical Spirit Sensitivity 5mv/grid to 500v/grid
Vertical resolution 8-bit (256 layers)
record length 512 samples per channel
trigger source Output 1, Input 2
Time base mode Normal, scrolling, single shot
Pre trigger Up to 10 grids
Measurement function AC, DC, AC+DC, peak value, peak to peak value, frequency, duty cycle, phase, pulse width, peak factor
Resistance, diode, on/off
resistance 500.0ω, 5.000kω, 50.00kω 500.0kω, 5.000mω, 30.00mω ± (0.6%+5 words)
Diode voltage ± (2%+5 words)
on-off Beehive<(30 Ω± 5 Ω)
Zui high current 0.5ma
capacitance 50.00nf, 500.0nf, 5.000μf, 50.00μf, 500.0μf ± (2%+10 words)
Temperature (optional) -100.0 ℃ to 400.0 ℃ -200.0 ° F to 800.0 ° F ± (0.5%+5 words)
Screen storage 10
Optoelectronic isolation RS232 interface
To the printer Support HP LaserJet ® deskjet ® Epson FX/LQ and Postsc Rippprinters using PAC91 printer converter
To PC Randomly including software
fluke view ® Power Quality Software
Hardware requirements Win31, Win95, Winnt environment, 4m hard disk space
Ordering Information
Random attachment Tl24 test line
Ac20 industrial crocodile clip ac85a large opening crocodile clip
TP2 and TP4 test probes
80i-500s 500a AC current clamp
Fluke View software
Pm9080/001 Optical RS232 Interface Adapter
BP120 rechargeable battery (Ni CAD, built-in)
PM8907 Line Voltage Adapter/Charger
BB120 Shielding Banana to Coaxial Adapter
C120 hard carrying box
User Manual
Application Guide
fluke view ® manual
F43B