The online phosphate analyzer is designed specifically for monitoring the discharge of pollutants from water sources, and is particularly suitable for online monitoring of phosphate at the outlets of factories, mines, and urban sewage treatment plants.
The analysis method of the instrument adopts the molybdenum antimony anti spectrophotometric method, which complies with the standard GB 11893-89 "Determination of Phosphate in Water Quality - Ammonium Molybdate spectrophotometric method" and belongs to the Class A method for phosphate monitoring, ensuring the accuracy and effectiveness of monitoring data.
Instrument performance:
Application field |
Online automatic measurement of phosphate in factories, mining enterprises, urban sewage and sewage treatment facilities |
measurement method |
Molybdenum antimony spectrophotometric method |
display screen |
5.6inchTFTcolor touch screen |
measurement range |
(0~2)mg/L;(0~10)mg/L;(0~50)mg/L;(0~250)mg/LCan be expanded according to user requirements |
measurement accuracy |
Quality control sample: ±10% |
zero drift |
±5% F.S. |
range drift |
±10% F.S. |
detection limit |
0.02mg/L |
measurement period |
Short measurement cycle20minute/Secondly; Can set online automatic monitoring cycle:0.6hour/time |
Reagent consumption |
0.8milliliter/time/Reagents; match withTP-BType reagent kit, can guarantee2Monthly full load total800Secondary measurement |
automatic calibration |
Automatic calibration can be set with small intervals1sky |
automatic cleaning |
The instrument automatically cleans after each measurement |
data storage |
50Storage of 10000 historical data records (full data records for over ten years) |
Interface output |
standard configurationRS232/RS485Two routes(4-20)mAOptional: Wireless IoT interface, micro printer, linkage control interface |
Usage environment:
external power supply |
(220±22)VAC;(50±2.5)Hz; Instrument power: daily average50WDuring peak hours<300W |
Ambient temperature and humidity |
0~40)oCRelative humidity≤90%,No condensation or dew formation |
External water sample |
Pressure:(-10~+2)KPa; Flow rate:(-1.0~0.5)m/s; Temperature:(5~40)oC |
Appearance dimensions:
Purchase configuration |
Instrument cabinet |
Core Analysis System |
Instrument size |
wide´tall´Thickness:420mm´600mm´250mm |
wide´tall´Thickness:390mm´290mm´210mm |
Net weight of the instrument |
20Kg |
6Kg |
Installation type |
Wall mounted installation or two in one cabinet installation |
Installed inside the instrument cabinet, detachable |