P15 Precision Conductivity and Dissolved Oxygen Dual Parameter Tester Instrument Purpose: Used for high-precision measurement of conductivity and dissolved oxygen in the laboratory. Suitable for industries such as biotechnology, aquaculture, power plants, and environmental protection.
P15 Precision Conductivity and Dissolved Oxygen Dual Parameter Tester Instrument Features:
7-inch color high-resolution (1024 * 600) capacitive touch screen, high sensitivity, complete parameter display, intuitive, easy to use, and simple to use. Compliant with international GLP requirements, it has intelligent functions such as automatic calibration, automatic temperature compensation, data storage, USB output, clock display, wireless printing, function settings, and self diagnostic information. The instrument is equipped with a bilingual operating system in both Chinese and English, with user-friendly operation design and instrument brightness The built-in large capacity storage can store 2000 sets of measurement data, which can be transferred through USB storage. The stored data can be opened with the general office software EXELL for easy editing by customers. The dissolved oxygen measurement mode has the functions of automatic temperature compensation, automatic salinity compensation, and automatic pressure compensation. The new structure of the polarographic dissolved oxygen electrode can simultaneously measure temperature, salinity, and dissolved oxygen. Equipped with a dedicated dissolved oxygen electrode calibration kit, electrode polarization only takes 3-5 minutes. Using a combination diaphragm cap, when replacing the sensitive membrane, simply tighten the new diaphragm cap onto the electrode, making it extremely convenient to use. Automatically recognize 8 types of conductivity standard solutions, with two series of standard solutions to choose from: European and American series and Chinese series. Adopting advanced conductivity measurement technology, the conductivity electrode with a constant K=1 only requires one point calibration to meet the measurement accuracy requirements of 0.5 μ S/cm~200 mS/cm. ● switchable conductivity TDS、 Salinity and electrical resistivity. TDS and salinity are calculated using polynomials to ensure full range conversion accuracy. The instrument is equipped with a Bluetooth module that supports wireless Bluetooth printing, making it convenient for users to operate. Optional support for wireless Bluetooth transmission of data to mobile phones and computers. Designed to meet the IP54 dust and splash protection rating.
P15 Precision Conductivity and Dissolved Oxygen Dual Parameter Tester Instrument Parameters:
Conductivity measurement range: Conductivity: (0.00~20.00)μS/cm (20.0~200.0)μS/cm
(200~2000)μS/cm (2.00~20.00)mS/cm
(20.0~200.0)mS/cm (200~2000)mS/cm(K=10)
Electrical resistivity: (0~100) M Ω· cm
TDS: (0 ~ 100)g/L
Salinity: (0 ~ 100)ppt
Resolution: 0.01/0.1/1 μ S/cm 0.01/0.1/1 mS/cm
Accuracy: Electrometer: ± 0.5% FS, Supporting: ± 1.0% FS
Temperature compensation range: (0~100) ℃ (automatic)
Reference temperature25 ℃, 20 ℃, and 18 ℃
Dissolved oxygen measurement range: (0 ~ 20.00)mg/L(ppm) (0 ~ 200.0)%
Resolution:0.1/0.01 mg/L(ppm) 1/0.1 %
Accuracy:Electricity meter: ± 0.10 mg/L, matching: ± 0.40 mg/L
Response time: ≤ 30 seconds (25 ℃, 90% response)
Residual current:≤0.1 mg/L
Temperature compensation range: (0-100) ℃ (automatic/manual)
salinityCompensation range: (0~45) ppt (automatic)
Pressure compensation range: (0~200) kPa
Automatic calibration: Air saturated with water; Water saturated with air
Electrode type: Polarographic
Data storage: 2000 sets
Power supply: 12V1A
Communication interface:Wireless Bluetooth (optional), USB2.0
Size and weight: 160 × 190 × 70mm/880g
Environmental temperature: 10~30 ℃
Environmental humidity: ≤ 80%
IP rating: IP54 dustproof and splash proof
|