Online Conductivity/TDS Analyzer InnoCon 6501C

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Online Conductivity/TDS Analyzer InnoCon 6501C Measurement Parameters: $r $n ● Conductivity/resistivity/TDS $r $n ● Temperature $r $n $r $n $r $n $r $n Applications: $r $n Wastewater, tap water, pure water, ultrapure water, boiler water, reverse osmosis, cooling towers, etc

Product Details

Online Conductivity/TDS Analyzer InnoCon 6501CProduct features:

● Large backlit LCD display, clear and easy to read 24/7

Economical and practical, standard 4-20Ma output

RS485 Modbus RTU communication

● 2-channel independent Hi/Lo alarm relay output

● Automatic temperature compensation function

● Wide power supply design, 90-260VAC

IP65 protection level, panel/pipeline installation InnoCon 6501C conductivity transmitter designed for water treatment and industrial process control, used with InnoSens 320/330 conductivity electrodes, can measure the conductivity, resistivity, TDS, and temperature of water samples. Chinese and English menus, automatic temperature compensation, standard 4-20mA/RS485 Modbus and relay alarm output.
Online Conductivity/TDS Analyzer InnoCon 6501CTechnical parameters:

Measuring range

parameter

measurement range

measurement accuracy

resolution

conductivity

0.00MS/cm – 2000mS/cm

±1%f.s.

0.01/0.1/1

resistivity

0.00 – 20.00MΩ*cm

±1%f.s.

0.01

TDS

0

±1%f.s.

1ppm

temperature

-10.0 - +130.0 °C

±0.2 C

0.1 C

Technical Specifications:

Measurement range: 0.00 μ S/cm – 2000mS/cm, 0.00 – 20.00M Ω* cm, -10.0~+120.0 ℃

Measurement accuracy: ± 1% f.s, ± 0.2 ℃

Resolution: 0.01/0.1/1 μ S/cm (mS/cm), 0.01 M Ω* cm, 0.1 ℃

Temperature compensation: PT1000, NTC30K

Analog output: isolated 4-20mA output, adjustable pH/ORP/temperature measurement values, Max 500 Ω

Contact output: 2-channel high/low alarm contact output, with hysteresis setting, 5A/250VAC

Digital Communication: RS485 Modbus RTU (9600, 8, N, 1)

Operating temperature: 0~70 ℃

Display: Backlit dot matrix LCD display

Language: Chinese/English

Power supply: 90-260VAC, 50/60Hz

Protection level: IP65

Installation method: panel type/pipe type

Dimensions: 108 × 108 × 145mm

Hole size: 92 × 92mm