| Resistance measurement range |
Range of 10 m Ω (maximum display 12.0000 m Ω, resolution 10 n Ω) to 1000 M Ω (maximum display 1200.0 M Ω, resolution 100 k Ω), with 12 level switching Range of 1000 m Ω (maximum display 1200.00 m Ω, resolution 10 μ Ω)~1000 Ω (maximum display 1200.00 Ω, resolution 10 m Ω), 4-speed switching Basic accuracy: ± 0.006% rdg. ± 0.001% fs |
| Measure current |
DC 1 A to 1 μ A or less, 1 mA to 5 μ A |
| Open circuit port voltage |
DC 20V (range 10 k Ω~), maximum 5.5 V (range~1000 Ω) When LP is turned on, the maximum DC voltage is 20 mV |
| temperature measurement |
-10.0-99.9 ° C, basic accuracy: ± 0.50 ° C (combined accuracy with temperature probe Z2001), -99.9-99.9 ° C (analog input) |
| measure speed |
FAST (2.2 ms), MED (50 Hz: 21 ms, 60 Hz: 18 ms), SLOW1 (102 ms), SLOW2 (202 ms) |
| function |
Temperature compensation, temperature conversion, cut-off voltage compensation (OVC), comparator (ABS/REF%), BIN, lock key (OFF/menu lock/overall lock), selection digit display function (7 bits/6 bits/5 bits), power frequency setting (AUTO/50Hz/60Hz), scaling, judgment sound setting, automatic hold, average value, statistical operation, panel save/read, D/A output |
| Multi channel adapter |
[* RM3545-02] Support Unit: Z3003 (2 additional units) |
| communication interface |
Choose one of GP-IB (* RM3545-01)/RS-232C/PRIMTER (RS-232C)/USB Remote function/communication monitoring function/data output function/storage (50 pieces) |
| power supply |
AC100~240 V, 50/60 Hz, Rated power: 40 VA |
| Volume and weight |
215W × 80H × 306.5D mm, [RM3545, RM3545-01] 2.5 kg, [RM3545-02] 3.2 kg |
| attachment |
Power cord x 1, clamp type test wire L2101 x 1, temperature probe Z2001 x 1, EXT I/O male connector x 1, user manual x 1, application software CD x 1, USB connection cable (A-B type) x 1, spare fuse x 1 |
|