Welcome Customer !

Membership

Help

Beijing Lijia United Technology Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products

Measurement of δ 18O and δ D of H2O isotopes using L2130-i isotope and gas concentration analyzer

NegotiableUpdate on 04/05
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Measurement of δ 18O and δ D isotopes of H2O

Product Details

The Picarro L2130-i isotope analyzer can achieve high-quality measurement of water stable isotopes, suitable for harsh applications such as paleoclimatology, hydrology, and oceanography. By using various Picarro peripheral devices, high-precision measurements of δ 18O and δ D can be performed on water samples taken from liquids, gases, and solids.


  • High precision measurement of δ 18O and δ D

  • Minimum drift: Calibrate once a day while achieving sub ppm accuracy measurement

  • Flexible measurement of water samples taken from liquids, gases, and solids

  • Small footprint, sturdy and durable design, intuitive user interface

For δ 18O/δ D, ensure that the accuracy of the liquid sample is 0.025/0.1 ‰ and the drift within 24 hours is 0.2/0.8 ‰. The water vapor measurement specifications include a measurement range of 1000 to 50000 ppm; For δ 18O (10/100 seconds) and δ D (10/100 seconds), ensure that the accuracy at 2500 ppm is 0.250/0.080 ‰ and 1.600/0.500 ‰, respectively, and the accuracy at 12500 ppm is 0.120/0.040 ‰ and 0.300/0.100 ‰, respectively.


L2130-i Technical Specifications

Picarro L2130-i Liquid Water Measurement Specification (equipped with A0211 and A0325) specifications Typical performance*
Standard mode quick mode
Accuracy (1 σ) Ensure:
δ18O – 0.025‰
δD – 0.1‰
δ18O – 0.010‰
δD – 0.05‰
δ18O – 0.015‰
δD – 0.05‰
Zero drift (24 hours) Ensure:
δ18O – 0.2‰
δD – 0.8‰
δ18O – 0.059‰
δD – 0.30‰
δ18O – 0.100‰
δD – 0.43‰
Sampling speed
(6 injections per sample; fast mode)
Next, each sample will be injected 10 times
54 minutes per sample/
27 samples per day
54 minutes per sample/
27 samples per day
29 minutes per sample/
50 samples per day
memory effect Ensure:
(After the third injection)
δ18O – 99%
δD – 98%
(After the third injection)
δ18O – 99%
δD – 98%
15 minutes later)
δ18O – 99%
δD – 98%
Total amount of solids dissolved in water < 200=''> Not Applicable Not Applicable

*Typical performance refers to the median of test results from multiple L2130-i analyzers manufactured continuously. Results can be provided upon request.


Picarro L2130-i Water Vapor Measurement Specification
measuring range 1000 to 50000 ppm
Ensure accuracy (1)σ)In
At a concentration of 2500 ppm
0.250/0.080 ‰, used for δ18O, 10/100 seconds
1.600/0.500 ‰, used for
δD, 10/100 seconds
Ensure accuracy (1)σ)In
At a concentration of 12500 ppm
0.120/0.040 ‰, used for δ18O, 10/100 seconds
0.300/0.100 ‰, used for
δD, 10/100 seconds
Measurement rate ~ 1 Hz


Picarro L2130-i analyzer specifications
measurement Cavity ring-down spectroscopy
temperature -10 to 45 ° C (water vapor sample); 10 to 35 ° C
(Liquid water sample and system operation); -10 to 50 ° C
(Storage conditions)
Sample pressure 300 to 1000 Torr (~40 to 133 kPa)
Sample flow rate At 760 pallets, 40 standard milliliters per minute, no need for filtration
Installation form Desktop or 19 inch rack mounted installation
Analyzer dimensions 17 inches wide x 7.5 inches high x 17 inches long
(43.2 cm x 19.1 cm x 43.2 cm)
Analyzer weight 45 pounds (20.4 kilograms)
Power requirements 90-240 volts AC, 50/60 Hz,<150 w=''>150>
(Analyzer), 80 W (external pump)
operating system Pre installed Picarro software for Windows 10 Professional