LEXT ™ OLS5100 3D laser scanningmicroscope
Laser microscope for material analysis
More intelligent workflow, faster experimental design
The OLS5100 laser microscope, suitable for fault analysis and materials engineering research, combines measurement accuracy and optical performance with intelligent tools, making the microscope more convenient to use. Simplify your workflow by quickly and efficiently measuring reliable data on sub micron shapes and surface roughness.
Ensure measurement accuracy
The LEXT microscope objective can provide highly accurate measurement data. When used in conjunction with the Smart Lens Advisor, reliable high-precision data can be obtained.

Guaranteed measurement accuracy*
Olympus' world-renowned optical devices can capture the true morphology of samples in the entire field of view with lower aberrations
The Smart Lens Advisor can also help you choose the appropriate lens for roughness measurement
*The information on this webpage, including the guaranteed accuracy, is based on the conditions set by Olympus.
Simplify the management of materials engineering and failure analysis experiments
The management of experimental conditions when testing new materials is complex, and the Smart Experiment Manager of OLS5100 laser microscope can automate and simplify this process through key steps such as creating experimental plans.
Speed up measurement tasks by 30%*
Scan plan automatically generated with data collection
No need to spend time transcribing data, the software automatically completes it for you

Automatic data input
The software will automatically add values to the experimental plan matrix to minimize the possibility of input errors. With just a few clicks of the mouse, experimental data can be exported to an Excel spreadsheet.

Simple experimental condition data management
By clicking on each cell in the experimental plan, the software can automatically generate a file name containing the evaluation criteria, making it convenient to save records. Each file contains associated images and data.

data collection
The microscope scans the sample according to the experimental plan generated by the software, which will not cause you to lose data or duplicate work.

Easily identify problems
Color charts can help you understand experimental data and verify the presence of data loss and errors. If there are any issues, you can identify and resolve them as early as possible.

Choose the appropriate objective lens
The Smart Lens Advisor can also help you choose the appropriate lens for roughness measurement applications. Simply select the appropriate lens and click start, and the Smart Lens Advisor will tell you the recommended level of the lens.

Easy data collection
Both experienced users and beginners can use the Smart Scan II feature to quickly and easily obtain data. Place the sample on the stage, press the "start" button, and the microscope will automatically complete the remaining operations.

| model | OLS5100-SAF | OLS5100-SMF | OLS5100-LAF | OLS5100-EAF | |
|---|---|---|---|---|---|
| Total magnification | 54x–17,280x |
||||
| field of view | 16 µm–5,120 µm |
||||
| Measuring principle | optical system | Reflective confocal laser scanning lasermicroscope Reflective confocal laser scanning laser DIC microscope color Color DIC |
|||
| Light receiving element | Laser: Photomultiplier tube (2 channels) Color: CMOS color camera |
||||
| Height measurement | display resolution | 0.5 nm | |||
| dynamic range | 16 bits | ||||
| Repeatability σ n-1*1 *2 *5 | 5X:0.45 μm, 10X:0.1 μm, 20X:0.03 μm, 50X:0.012 μm, 100X:0.012 μm | ||||
| accuracy*1 *3 *5 | 0.15+L/100 μ m (L: measured length [μ m]) | ||||
| Accuracy of stitched images*1 *3 *5 | 10X: 5.0+L/100 μ m, 20X or higher magnification 1.0+L/100 μ m (L: splicing length [μ m]) | ||||
| Measurement noise (Sq noise)*1 *4 *5 | 1 nm (typical value) | ||||
| width measurement | display resolution | 1 nm | |||
| Repeatability 3 σn-1 *1 *2 *5 | 5X:0.4 μm, 10X:0.2 μm, 20X:0.05 μm, 50X:0.04 μm, 100X:0.02 μm |
||||
| accuracy*1 *3 *5 | Measurement value+/-1.5% | ||||
| Accuracy of stitched images*1 *3 *5 | 10X:24+0.5L μm, 20X:15+0.5L μm, 50X:9+0.5L μm, 100X:7+0.5L μm (L: 拼接长度 [mm]) | ||||
| Number of measurement points during a single measurement | 4096 × 4096 pixels |
||||
| number of measurement points | 36 million pixels |
||||
| XY stage configuration | Length measurement module | • | Not applicable | Not applicable | • |
| scope of work | 100 × 100 mm (3.9 × 3.9 in.) electric |
100 × 100 mm (3.9 × 3.9 in.) manual |
300 × 300 mm (11.8 × 11.8 in.) electric |
100 × 100 mm (3.9 × 3.9 in.) electric |
|
| Sample height | 100 × 100 mm (3.9 × 3.9 in.) |
30 mm (1.2 in.) |
30 mm (1.2 in.) |
210 mm (8.3 in.) |
|
| laser light source | wavelength | 405 nm | |||
| output | 0.95 mW | ||||
| Laser Class | Level 2 (IEC60825-1:2007, IEC60825-1:2014) | ||||
| Color light source | white LED | ||||
| Electrical power | 240 W | 240 W | 278 W | 240 W | |
| quality | microscope body | About 31 kg | About 32 kg | About 50 kg | About 43 kg |
| control box | About 12 kg | ||||
*Provide a guarantee when used in a constant temperature and humidity environment as specified in ISO554 (1976) and JIS Z-8703 (1983) (temperature: 20 ° C ± 1 ° C)? C, Humidity: 50% ± 10%), as specified in ISO554 (1976) and JIS Z-8703 (1983).
*When measuring with MPLAPON LEXT series objective lenses, the magnification is 20x or higher.
*When measuring with a dedicated LEXT objective lens.
*Typical values when measuring with MPLAPON100XLEXT objective lens.
*5 guarantees based on the Olympus certification system.
**The OS license for Windows 10 has been certified by Olympus' microscope controller. Therefore, accepting the Microsoft License Terms means that you agree to these terms. For Microsoft's license terms, please refer to the following content.
/en-us/Useterms/Retail/Windows/10/UseTerms_Retail_Windows_10_english.htm
| series | model | Numerical aperture (NA) | Working distance (WD) (mm) |
|---|---|---|---|
| UIS2? objective lens | MPLFLN2.5x | 0.08 | 10.7 |
| MPLFLN5x | 0.15 | 20 | |
| Specialized LEXT objective lens (10X) | MPLFLN10xLEXT | 0.3 | 10.4 |