LI-7500DS Open Circuit CO2/H2O Intelligent Analysis System

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The LI-7500 series (including LI-7500, LI-7500A, LI-7500RS, LI-7500DS) open circuit CO2/H2O gas analyzer has been widely used in the field of vorticity covariance carbon flux monitoring. More than 90% of the flux observation stations of the carbon flux observation network (FLUXNET) use the LI-7500 series CO2/H2O gas analyzer.

Product Details

LI-7500DS is a new generation of digital open circuit CO2/H2The O intelligent analysis system is designed for rapid measurement of CO in the atmospheric environment2And H2Designed by O,The low power consumption of up to 4W and almost maintenance free characteristics make it more suitable for stable measurements at field observation stations.

Main Features

LI-7500DS

Low power consumption:Ensure that it can perform positioning measurements on field stations

LI-7500DS

*Continuous measurement:It can sample and measure airflow in all directions, allowing for multiple installation options and ensuring continuous data measurement

LI-7500DS

Record complete vorticity covariance dataIncluding wind speed data collected by ultrasonic anemometers and auxiliary meteorological, radiation, and soil data collected by LI-COR Biomet system


working principle

LI-7500DS uses non dispersive infrared spectroscopy to measure CO in the air2And H2The density of O. Infrared rays pass through a temperature controlled filter and then through an open sampling space to reach the infrared detector. CO in the sampling space of infrared radiation with specific wavelengths2And H2O absorption, obtaining CO by calculating the degree of infrared attenuation in this band2And H2The gas concentration of O.

New features of the instrument

1. Accurate temperature control of the light source and detector ensures measurement stability even in large environmental temperature fluctuations.
2. The precise optical components greatly reduce its sensitivity to pollutants such as pollen, dust, etc. in natural environments.
3. Design of sapphire scratch resistant lenses with special light waves.
4. High performance controllable temperature brushless chopper ensures reliable operation.

built-inSmartFlux®system

SmartFlux®The system will integrate EddyPro®The functionality of the software is brought to your field site, like a powerful field computer that can perform real-time and complete vorticity flux calculations. The calculation results are consistent with your use of EddyPro in the office®The results obtained from processing raw data are the same, but more timely. SmartFlux®The system is FluxSuiteTMThe hardware foundation of the data online monitoring and management system.

technical indicators


LI-7500DS

CO2measurement
·Calibration range: 0~3000 µ mol/mol
·Accuracy:<1%
·Zero drift (per ℃):
Typical ± 0.1 µ mol/mol
o zui 大 ±0.3µmol/mol
·RMS noise/resolution (typical @ 370 µ mol/mol CO)2):
o @5 Hz:0.08µmol/mol
o @10 Hz:0.11µmol/mol
o @20 Hz:0.16µmol/mol
·Gain drift (% of reading per ℃ @ 370 µ mol/mol):
Typical ± 0.02%
o zui 大 ±0.1%
·Regarding H2Sensitivity of O (mol CO)2/mol H2O):
Typical ± 2.00E-05
o zui 大 ±4.00E-05

H2Omeasurement
·Calibration range: 0~60 mmol/mol
·Accuracy:<1%
·Zero drift (per ℃):
Typical ± 0.03 mmol/mol
o zui 大 ±0.05 mmol/mol
·RMS noise/resolution (typical @ 10mmol/molH)2O):
o @5 Hz:0.0034 mmol/mol
o @10 Hz:0.0047 mmol/mol
o @20 Hz:0.0067 mmol/mol
·Gain drift (% of reading per ℃ @ 20 mmol/mol):
Typical ± 0.15%
o zui 大 ±0.30%
·Regarding CO2Sensitivity (mol H)2O/mol CO2):
Typical ± 0.02
o zui 大 ±0.05

comprehensive
·Basic gas sampling frequency: 150Hz bandwidth: 5,10 or 20Hz;
·Analyzer: Open circuit non dispersive infrared solid-state detector
·Detector: Thermoelectric cooled lead selenide optical path length 12.5cm
·Data communication: Ethernet
·Air temperature thermistor: 10K ohm@25 ℃ thermistor
Measurement range: -40~70 ℃
Accuracy: ± 0.25 ℃ @ -20~70 ℃
Resolution: 0.003 ℃ @ 25 ℃
·Pressure sensor
Measurement range: 20-110 kPa
Accuracy: ± 0.4 kPa @ 50~110 kPa
Resolution: 0.006kPa
·Working temperature range: -25 to 50 ℃ (can be expanded to -40~50 ℃ as needed)
·Relative humidity range: 0-95% (non condensing) Waterproof level IEC IP65
·Power supply: 10.5~30 VDC
·Power consumption: Typical 4W @ 25 ℃, maximum 8W @ -25 to 50 ° C during operation
·User Interface: Windows®platform

Analyzer probe
·Size: Diameter 6.5 cm, Length 30 cm
·Weight: 0.67 kg; base connected 1.3kg; cable length 200cm

DSIbox
·Size:
13.24cm × 14.64cm × 6.24 cm(H×W×D)
·Weight: 0.93kg