GC9800

NegotiableUpdate on 03/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
1. Column box -8 ℃ to 400 ℃ above room temperature; 2. Vaporization chamber, detector - above room temperature 20 ℃~380 ℃ 3, switching valve - above room temperature 20 ℃~160 ℃. 4. Temperature stability of 70 ℃ column box:<0.5%. 5. Temperature display accuracy: 0.1 ℃. 6. Column box program heating: 8 stages, rate 0-30 ℃/min (minimum increment 0.1 ℃/min). 7. Column box program heating (50 ℃ -10 ℃/min -250 ℃) repeatability: ≤ 1%.
Product Details

1、 Temperature control:

Temperature control range:

1. Column box -8 ℃ to 400 ℃ above room temperature;

2. Vaporization chamber, detector - above room temperature 20 ℃~380 ℃;

3. Switching valve - above room temperature 20 ℃~160 ℃;

4. Temperature stability of 70 ℃ column box: < 0.5%;

5. Temperature display accuracy: 0.1 ℃;

6. Column box program heating: 8 stages, rate 0-30 ℃/min (minimum increment 0.1 ℃/min);

7. Column box program heating (50 ℃ -10 ℃/min -250 ℃) repeatability: ≤ 1%;

8. Cooling rate of column box: ≤ 9 minutes (ambient temperature<27 ℃, vaporization temperature: 160 ℃, detection temperature: 180 ℃, column box temperature drops from 105 ℃ to 35 ℃).

2、 Detection sensitivity:

1. FID detection limit Mt ≤ 5 × 10-12g/s

2. 1 µ l of nC16/iso octane standard solution with a concentration of 144.4ng/µ l, the peak area of nC16 after injection should be ≥ 26000 µ V.s

3、 FID baseline stability:

(1) Specialized filling column, 135 ℃ constant temperature;

Baseline noise<10 × 10-14A

Baseline drift<6 × 10-13A/30min

(2) Specialized capillary column, programmed heating from 35 ℃ to 105 ℃, FID high resistance 1010 Ω;

Baseline noise<50 µ V

Baseline drift<1000 µ mV/30min

Startup time: 1 hour (from startup to sample analysis).