When coupled with peripheral devices such as the overall combustion module (A0201) for sample preparation, the Picarro G2121-i isotope and gas concentration analyzer is capable of measuring δ 13C and CO2 gas concentration in high concentrations of carbon dioxide (CO2) (2000-4000 ppm). The accuracy of measuring δ 13C in CO2 is better than 0.4 ‰ (typically 0.2-0.3 ‰).

Used in conjunction with the overall combustion module and AutoMate FX water dissolved inorganic carbon (DIC) sample preparation peripheral equipment
High precision, high accuracy, minimum drift
No consumables required for on-site and laboratory deployment
Durable and sturdy structure, insensitive to environmental temperature changes
This analyzer is designed with a precise chamber temperature and pressure control system, which can ensure accurate measurement for a long time and achieve minimal calibration. This analyzer has a very sturdy structure, is insensitive to constantly changing environmental temperatures, has almost no drift and maintenance, and requires almost no consumables. This analyzer can be easily transported between stations and can be set up and data collected within 60 minutes.
| Picarro G2121-i Target Performance Specifications | |
| Gas components | Accuracy (5-minute sample peak average) |
| In CO2δ13C | The deviation between samples is less than 0.4 ‰ (typically 0.2-0.3 ‰) |
| Range of 12CO2 concentration | 2000–4000 ppm |
| Picarro G2121-i System Specifications | |
| measurement | Cavity ring down spectroscopy |
| range | CO2:2000–4000 ppm |
| Measurement Interval | About 10 seconds |
| sample temperature | -10 to 45 ℃ |
| Sample flow rate | Less than 0.1 standard liters per minute (slm) at 760 pallets, no filtration required |
| Sample pressure | 300 to 1000 Torr (40 to 133 kPa) |
| Sample humidity | Relative humidity (RH) less than 99%, no need to dry under non condensing conditions at 40 ℃ |
| ambient temperature | 10 to 35 ℃ (during operation) -10 to 50 ℃ (during storage) |
| ambient humidity | Relative humidity (RH) less than 99%, under non condensing conditions |
| Other gases measured | H2O |
| attachment | Pump (external), keyboard, mouse, LCD monitor (optional) |
| data output | RS-232、 Ethernet, analog (optional) 4-20 mA/-10-10V |
| Air inlet connector | ¼ inch Swagelok ® |
| External dimensions | Analyzer: 17 inches wide x 7 inches high x 17.6 inches long (43.2 x 17.8 x 44.6 centimeters), excluding 0.5 inch feet External pump: 5.6 inches wide x 6.4 inches high x 11.9 inches long (14.3 x 16.3 x 30.3 centimeters) |
| install | Workbench |
| weight | 60.4 pounds (27.4 kilograms), including external pump |
| Power requirements | 90-120 volts AC, 50/60 Hz; 220 volts AC, 50 Hz. Less than 300 watts. |