Welcome Customer !

Membership

Help

Houpu Scientific Instrument Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products

Fully automatic oil acid value tester

NegotiableUpdate on 01/22
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The TACH-PI-001 fully automatic oil acid value tester automatically performs operations such as sample preheating, sample selection, neutralization liquid filling, addition of anhydrous ethanol, heating, titration blank, injection, heating, neutralization titration, titration endpoint determination, result calculation, waste disposal, pipeline cleaning, and printing of test results;

Product Details

1、 TACH-PI-001 typeFully automatic oil acid value testerProduct Features:

  • This instrument adopts a Weilun Tong 10 inch color touch screen and has a good human-machine operation interface;
  • The measurement interface of this instrument has an intuitive measurement flowchart and displays the measurement progress, allowing operators to better understand the measurement status of the instrument;
  • The instrument is easy to operate and adopts a one click measurement mode. Simply retrieve the sample with the matching sampling device and connect it to the corresponding position on the instrument. The instrument automatically performs sample preheating, sample selection, neutralization liquid filling, addition of anhydrous ethanol, heating, titration blank, injection, heating, neutralization titration, titration endpoint judgment, result calculation, waste disposal, pipeline cleaning, and printing of test results;
  • The measurement process of this instrument is automated and does not require human operation, greatly reducing errors caused by human factors;
  • This instrument adopts a self-developed safe, durable, reliable, and accurate automatic sampling, injection, neutralization titration, and waste disposal system to ensure the continuous and normal operation of the instrument;
  • This instrument has a self calibration system that can calibrate the concentration of the neutralizing solution, automatically calibrate and deduct the acidity value of anhydrous ethanol (blank value), and calibrate the measurement of the neutralizing solution, greatly reducing measurement errors caused by various factors and improving measurement accuracy;
  • This instrument uses different titration endpoint determination methods for different samples to ensure the accuracy and reliability of experimental data;
  • *The design of the anhydrous ethanol and neutralizing liquid delivery pipeline ensures that the reagents do not come into contact with air during use, avoiding the influence of solvent evaporation and acidic gases in the air on the experiment, and reducing the harm to the staff's body;
  • Users can choose the method of determining the number of samples to be placed simultaneously and the endpoint of titration according to their own needs;
  • The instrument automatically discharges waste and cleans related pipelines after the measurement is completed, thereby reducing people's workload;
  • This instrument can store 500 measurement data and has a historical data query function. It can record and print each required data as needed, and can be queried at any time;
  • The instrument has the characteristics of resistance to voltage fluctuations, environmental electromagnetic interference, and strong electromagnetic compatibility (EMC). The appearance adopts a streamlined design, with a delicate and beautiful structure, and comfortable and convenient operation.

IIFully automatic oil acid value testerTechnical Specifications:

  • Measurement range: Acid value: 0.001-4mg KOH/g Acid value: 0.01-1000mg KOH/100mL
  • Execution standard: GB/T258-2016 Determination of acidity of light petroleum products

GB/T264-1983 Determination of Acid Value of Petroleum Products

GB/T28552-2012 Determination of oleic acid value of transformer oil and turbine oil (BTB method)

  • 1. Accuracy: Acid value: between 0.001~0.100 mg KOH/g, ≤± 0.001 mg KOH/g

Between 1.000 and 4.000 mg KOH/g, ≤± 0.01 mg KOH/g

Acidity: between 0.01~10.00mg KOH/100mL, ≤± 0.02mg KOH/100mL;

Between 10.00 and 100.00mg KOH/100mL, with a tolerance of ≤± 0.2 mg KOH/100mL

2. Repeatability: Acid value: between 0.001~0.100 mg KOH/g,<0.002 mg KOH/g;

Between 1.000 and 4.000 mg KOH/g,<0.02 mg KOH/g

Acidity: between 0.01 and 10 mg KOH/100mL,<0.02 mg KOH/100mL;

Between 10.00-100.00 mg KOH/100mL,<0.2 mg KOH/100mL

  • Resolution: Acid value: 0.001mg KOH/g Acid value: 0.01mg KOH/100mL
  • Display mode: Weilun Tong 10 inch color touch screen
  • Temperature sensor: Platinum resistance (imported * PT100)
  • Temperature control method: microcontroller control and PID self-tuning temperature control technology
  • Temperature control accuracy: 0.1 ℃
  • Measurement method: combination of reflux method and neutralization titration method;
  • Endpoint detection method: A: combination of indicator and photoelectric detection B: pH electrode detection
  • Measurement method: Imported high-precision flow adjustable micro ceramic metering pump
  • Sample quantity: 2, 4, 6, 8, 10, 12 (optional)
  • Printing method: The instrument is equipped with a built-in thermal printer
  • Clock accuracy: ± 1 minute/month
  • Continuous data retention time in case of power failure:>5 years
  • Working power supply: 220V ± 10% 50 ± 1 Hz
  • Working environment: Environmental temperature: 0-60 ℃
  • Environmental humidity:<85%