Economically expand the processing capacity of your laboratory using ECM8 electrochemical multiplexer. The experiment can be multiplexed in sequence on up to 8 channels. ECM8 is an ideal instrument for corrosion inhibitor testing, paint/coating EIS evaluation, field strength probe monitoring in corrosion testing, and chemical sensor development.
ECM8 ™ The combination of electrochemical multiplexer and Gamry potentiostat can convert a single potentiostat into a powerful instrument suitable for sequential multi-channel operation. In this way, you can measure in sequence on up to 8 electrochemical cells. ECM8 is very suitable for automating repetitive experiments and can greatly increase the processing capacity of long-term experiments (which can collect data regularly). And any one of the eight multiplexer channels, when not involved in electrochemical experiments, will also be combined with a "local" potentiostat to polarize the sample. ECM8 cannot be used for multi-channel synchronous experiments across multiple electrochemical cells. For information on multi-channel synchronous experiments, please refer to the MultEchem electrochemical system.Overview
application
Standard features
Channel characteristics
pattern
Activity, open circuit, short circuit, local
Channel switching time
<4 msec
Excluding communication time
Electrochemical cell current
1 ampere (maximum) after switching
Input voltage (during operation)
± 12V comparison chassis (excluding counters)
Input voltage (lossless)
± 24V comparison chassis
Counter electrode voltage
± 120V comparison chassis
Channel insulation
Current leakage
<2 nA (maximum) Any electrode from any source in the active channel
impedance
>500 MΩ <20 pF
ECM8 grounding
Power input
Input voltage range
100,120,220,240 V rms
Allowable voltage variation
± 10% within various ranges
power
<50 VA
Local potentiostat
(ECM8 only)
Constant current output - current
± 20 mA -500 Ω load
Constant current output - voltage
± 11 Volts -1k Ω load
Range of applied voltage difference
± 5.12 V working voltage compared to reference voltage
resolution
2.5 mV/bit
Zero bias error
<4 mV typical V output
<10 mV maximum 10 k Ω load
Gain error
<0.2% -10 k Ω load
Reference input current
<50 pA
Note: Voltage application error
RS-232 通信
Baud rate
300, 600, 1200, 2400, 4800, 9600, 19200
Set through internal DIP switch
default value
9600- Factory settings
handshake
Hardware exchanges signals through RTS and CTS
environment
Operating temperature
0 to 45° C
humidity
0 to 90% - No condensation