Welcome Customer !

Membership

Help

Jiangsu Jiamei Instrument Manufacturing Co., Ltd
Custom manufacturer

Main Products:

mechb2b>Products

Jiangsu Jiamei Instrument Manufacturing Co., Ltd

  • E-mail

    2891149@163.com

  • Phone

    13921015010

  • Address

    No. 249 Liangchang Middle Road, Jintan District, Changzhou City, Jiangsu Province

Contact Now

ETW-2 type air microbial sampler

NegotiableUpdate on 03/03
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The ETW-2 air microbial sampler adopts the Anderson sampler of *, with a large sampling flow rate, good stability, AC power supply, convenient use, compact size, and light weight, making it a powerful tool for on-site microbial sampling.
Product Details

ETW-2 type air microbial samplerIt is a small portable sampler. A new type of instrument that can be used for air microbiological monitoring and experimental research in medical and health, food industry, fermentation industry, pharmaceutical industry, agriculture and animal husbandry, as well as related research departments.

Microbial Air Monitoring SystemsCompared with the same type, it has the characteristics of wide applicability, small size, flexibility, and relatively complete components; The capture rate of microorganisms in the air is equivalent to that of a standard sampler.

The timer control is designed with microcomputers, and the error accuracy can reach over 0.5%.

Standard quality flow meter is a guaranteeETW-2 type air microbial samplerThe components of system sampling accuracy *.

Usage of air microbial sampler:

● Indoor air quality ● Hospital environment ● Bacterial growth

● Medicinal products ● Food processing plants ● Efficiency research on filters and cleanrooms

ETW-2Parameters of the Type Air Microbial Sampler:

Sampling pump constant current sampling, sampling flow rate5Adjustable up to 30 L/min

  • The flow meter has a range of 5-30 liters per minute. The adjustable flow device can be freely adjusted from 5 to 30 liters per minute. Usually using 20 liters per minute.
  • Aerosol made from Bacillus cereus spores. The capture rate is determined to be over 95% in a standard aerosol chamber.
  • The timing range of the timer is adjustable from 1-999 minutes, with an error not exceeding ± 5 ‰ (0.5 ‰).
  • The power supply method is AC 220V ± 10%, 50Hz.
  • Power consumption ≤ 45W.
  • Noise ≤ 68ab.
  • Volume: 290 × 220 × 200mm3.
  • Weight: 7Kg