- Phone
-
Address
B108, No. 352 Outer Ring Road, Minhang District, Shanghai (201199)
Shanghai Yifeng Precision Instrument Co., Ltd
B108, No. 352 Outer Ring Road, Minhang District, Shanghai (201199)
| Measurement project | |
| Measurement project | Measurable roundness, coaxiality, concentricity, runout, perpendicularity, parallelism, flatness, surface waviness (Wc, Wp, Wv, Wt, Wa, Wq), frequency spectrum analysis, wave height analysis, etc. of circular workpieces Parameters such as roundness, runout, single piece runout, adjacent piece runout, and intermittent difference of the motor commutator |
| measuring range | |
| Maximum measurement diameter | φ300mm |
| Maximum measurement height | 350mm |
| Capacity | 25Kg |
| Spindle Precision | |
| Radial error of spindle H=to the height of the workbench |
±(0.0125+3H/10000)μm |
| Axial error of spindle X=distance to the center of the spindle |
±(0.0125+4X/10000)μm |
| workbenches | |
| Table diameter | φ180mm |
| adjustment range | Adjust the eccentricity by ± 2mm; Adjust the level ± 1 ° |
| Rotation diameter | φ300mm |
| rotational speed | 0-12 revolutions per minute |
| Horizontal arm (horizontal guide rail) | |
| Horizontal movement distance | 160mm |
| movement speed | 0.5-6mm/S or manual |
| sensor | |
| range | 500 μ m (radius difference) |
| resolution | 0.005μm |
| Measuring needle shape | 2mm gemstone ball measuring head (with more diameters to choose from) |
| dynamometry | 1-12g |
| data acquisition | |
| data acquisition | 7200 points per week (internationally recognized) |
| Magnification factor | |
| Magnification factor | 200000 times automatic |
| appraising method | |
| appraising method | LSC (Least Squares Method), MZC (Minimum Region Method) MIC (maximum inscribed circle method), MCC (minimum circumscribed circle method) |
| Filter value | |
| Filter value | 1-500、1-150、1-50、1-15、3-16、3-17、15-500、17-100 upr |
| Data analysis and processing | |
| Spectrum analysis | Analyze the amplitude of different frequency components |
| Abnormal data removal | Manually or automatically remove abnormal data such as burrs, holes, and bumps |
| file management | Measurement data is automatically saved and can be deleted |
| Data printing | Single or multiple measurement results can be printed on the same page in the same printing mode |
| INSTALLATION REQUIREMENTS | |
| power supply | 220-240V AC, 50HZ, Separate ground wire; Maximum power consumption 450W |
| compressed air | Pure compressed air: pressure 0.4MP, consumption 0.1m3/min (air supply 0.6MP, 0.25m3/min or above) |
| Environmental requirements | Working temperature: 10 ° -30 ° C (normal temperature); Humidity 20-80% RH (no condensation), good grounding, no obvious vibration source |