Name: Electric constant temperature blast drying oven
Model: DGF3010B
Brand: Chongqing Sida
Chongqing Sida Electric Thermal Constant Temperature Blast Drying Oven 300 ℃ High Temperature OvenProduct Usage
The electric constant temperature blast drying oven is suitable for scientific research and production units such as electricians, electronics, instruments, materials, and pharmaceuticals. It is mainly used for drying, baking, heat treatment, disinfection, insulation, and other temperature tests of items.
Chongqing Sida Electric Thermal Constant Temperature Blast Drying Oven 300 ℃ High Temperature OvenPerformance characteristics:
1. Complete variety and specifications to meet the needs of different customers;
2. Assemble the structure with blocks, spray paint the shell in trendy colors, and create an elegant and beautiful appearance;
3. Forced convection duct structure form;
4. PID regulation, digital display instrument;
5. Equipped with independent studio over temperature protection and short circuit protection;
Technical parameters of electric constant temperature blast drying oven:
Model number |
Dimensions cm |
Studio size cm |
Volume L |
Temperature range ℃ |
temperature Volatility ℃ |
temperature uniformity |
install Power KW |
DGF3006 |
100×130×128 |
80×80×100 |
640 |
R.T+20~300 |
±0.5 |
±2.5% |
5.5 |
DGF3006B |
100×130×128 |
80×80×100 |
640 |
R.T+20~300 |
±0.5 |
±2.5% |
5.5 |
DGF3010B |
|
100*100*100 |
1000 |
R.T+20~300 |
±0.5 |
±2.5% |
7.2 |
B is a stainless steel inner liner | |||||||
Daily maintenance and general troubleshooting:
1. Daily maintenance: ⑴ The equipment should be kept clean. It is strictly prohibited to use corrosive cleaning agents to wipe equipment.
⑵ Do not violently strike or collide with equipment.
If the equipment is not in use after shutdown, the power plug should be unplugged and the equipment should be powered on regularly.
The equipment should be regularly (usually once a month) inspected by professional personnel and maintenance records should be kept.
If the equipment malfunctions, a fault record should be kept.
2. Fault handling:
Fault phenomenon |
Possible reasons |
solution |
Unable to heat up |
Incorrect over protection setting Heating wire breakage The thyristor is broken The temperature control instrument is broken |
Reset the overtemperature value Replace the heating wire Replace the thyristor replace the instrument panel |
Display jumping |
1 . The temperature control instrument is broken 2. Poor contact of the resistor |
replace the instrument panel Check the wiring of the resistor |
Temperature cannot be controlled |
1 . The temperature control instrument is broken 2. The thyristor is broken |
1. Replace the instrument panel 2. Replace the thyristor |
Slow heating up |
The heating wire is partially damaged Thyristor semiconductor conduction The temperature control instrument is broken |
Replace the heating wire Replace the thyristor replace the instrument panel |