Its purpose is to obtain temperature faster and more accurately for VICTOR25, 11+, 14+(applicable to version 1.0)
Its purpose is to obtain temperature faster and more accurately for VICTOR25, 11+, 14+(applicable to version 1.0).
These calibrators already have a temperature sensor inside, but if the external environment temperature changes quickly, such as from indoors to outdoors, the internal temperature sensor takes a long time to sense this change. If you start working outdoors immediately, there will be significant compensation errors. To accurately compensate, you will need to wait for a long time, which will reduce work efficiency. By using the CALCT temperature probe, you can overcome the above problems because it has a large metal plate closely attached to the internal temperature sensor, so it can quickly measure changes in ambient temperature. At the same time, its accuracy is ± 0.2 ℃, which is higher than the internal sensor's ± 0.5 ℃.
Model: A000019
Applicable products: Our company's VC25, VC11+, VC14+process calibrators
Function: Faster and more accurate measurement of ambient temperature and obtaining cold end compensation temperature;
Accuracy: ± 0.2 degrees
Features: MINI-USB interface (connected to instrument panel)
Weight: Approximately 17g