The ozone aging test chamber is one of Wuhe's flagship products, with free equipment delivery, installation and commissioning, and one-year after-sales service.
1、 Working principle
Static tensile test
1Beijing, Shanghai, Guangzhou Ozone Aging Test Chamber ManufacturerThe standard stipulates that the two ends of the sample should be clamped on a fixture, and a stretching rate of 20% should be selected. The fixture should be installed on a rotating sample holder,
2. The sample rack speed is between (20-25) mm/S, the temperature is (40 ± 2) ℃, the relative humidity is not greater than 65%, and the specified ozone concentration is used for testing.
3Beijing, Shanghai, Guangzhou Ozone Aging Test Chamber ManufacturerCheck the cracking condition of the sample after the specified time. The general selection of ozone concentration is (50 ± 5) × 10-8.
Dynamic tensile test
1. The standard stipulates that the sample fixture can automatically stretch and reset the sample, with a stretching frequency of 0.5 ± 0.025Hz. The two ends of the sample are clamped on the fixture, and the sample is cyclically stretched from 0 to a higher elongation rate. The fixture is installed on a rotating sample holder.
2. The sample rack speed is between (20-25) mm/S, at a temperature of (40 ± 2) ℃, relative humidity not exceeding 65%, and at the specified ozone concentration.
3. The fixture undergoes cyclic stretching and resetting, and rotates along with the sample holder for testing.
4. Check the cracking condition of the sample after the specified time. The general selection of ozone concentration is (50 ± 5) × 10-8.
5. Dynamic tensile testing is divided into continuous dynamic tensile testing and intermittent dynamic tensile testing.
2、 Control system
The temperature, humidity and concentration control part of the equipment adopts the PLC programmable controller imported from Taiwan, China, which is simple in programming, stable in temperature, humidity and concentration control, and large screen touch screen display and operation. The main electrical components of the equipment, such as circuit breakers, AC contactors, and small relays, are all made of Schneider brand, while other electrical accessories are made in China.
The time controller of the equipment controls the time required for the experiment and is easy to operate. When the device reaches the set time of operation, it automatically stops the test and enters a state of equilibrium to cool down.
3、 Refer to and implement standards
GB/T 13642-92 Vulcanized rubber - Ozone aging resistance test - Dynamic tensile test method;
GB/T 7762-2003 Rubber, vulcanized or thermoplastic - Static tensile test method for resistance to ozone cracking;
GB/T 2951.1-2951.0-1997 General Test Methods for Insulation and Sheath Materials of Cables;
And the test methods specified in other relevant standards.
4、 Ozone generator
1. Input voltage: 220V
2. Current: 300mA
3. Manual and automatic adjustment of concentration
4. Concentration range: 0-100%
5. Analog input voltage: 0-10VDC
6. Analog output voltage: 0-1VDC
7. High concentration of ozone: 120gO3/m3
8. Ozone production (oxygen intake, flow rate of 100L/h) 3.5gO3/h
9. Electrode cooling method: forced air cooling
10. Operating pressure: less than 0.2Mpa
11. Multiple groups can be used in parallel to meet design requirements.
5、 Main technical parameters
1. 型号 工作室尺寸 (D×W×H) 外形尺寸 (D×W×H) 功率
QL—100 400×500×500 1000×900×1400 2.5KW
QL—150 500×500×600 1100×900×1500 2.5KW
QL—225 500×600×750 1100×950×1600 3.5KW
QL—500 700×800×900 1300×1200×1700 6.0KW
QL—800 800×1000×1000 1400×1400×1850 7.5KW
QL—010 1000×1000×1000 1600×1400×1850 9.5KW
2. Temperature range: RT+10 ℃~50 ℃
3. Temperature uniformity: ≤ 2 ℃
4. Temperature fluctuation: ± 0.5 ℃/H
5. Temperature deviation: ≤± 2 ℃
6. Time setting range: 0-999 hours
7. Ozone concentration: 0-500PPHM, 0-1000PPHM
8. UV detection probe, long lifespan, high accuracy
9. Sample rack speed: (20-25) mmS (adjustable)
10. The time for the same specimen to rotate once is 8-12 minutes
11. Odorous oxidation air flow rate: 8-16mmS
12. Fixture elongation: 5% to 35%
13. Power supply voltage: AC 380V
6、 Structure of ozone aging test chamber
1. The ozone aging test equipment is made of high-quality materials, and the surface of the shell is treated with spray coating, which is beautiful and smooth.
2. The inner liner is designed to be a sealed space without direct sunlight during operation, where samples can be placed for aging tests. It can maintain a constant temperature difference of ± 2 ℃ during testing and is made of imported high-quality stainless steel plates.
3. The indoor sample racks, fixtures, conduits, and other accessories are made of materials such as stainless steel and aluminum that are not easily decomposed by ozone and affect ozone concentration.
4. The lifting height, frequency, and amplitude of the dynamic sample holder can be adjusted. The sample holder can rotate 360 degrees with adjustable speed.
5. Set a transparent window at a reasonable position on the box door to observe changes in the indoor sample.
6. The observation window adopts multi-layer hollow tempered glass, which has the advantages of transparency, heat insulation, and is not prone to steam frosting.
7. The interlayer between the inner and outer shells of the ozone aging test equipment is filled with high-density glass fiber cotton to ensure that the box will not experience excessive temperature fluctuations due to temperature conduction, and the temperature of the sample will remain constant throughout the test process.
8. The ozone aging test equipment uses a long axis fan motor for stirring and a stainless steel multi blade impeller that is resistant to high and low temperatures to achieve strong convection and vertical diffusion circulation, ensuring uniform and stable temperature in the laboratory.
9. A double-layer high-temperature resistant high tensile sealing strip is used between the door and the box to ensure the sealing of the testing area. Using non reactive door handles makes operation easier.
10. Set one or more testing holes in a reasonable position on the left side of the box, which can be connected to external testing power or signal lines. The arrangement of hole diameter or number of holes should be indicated by the user.
11. High quality fixed PU movable wheels are installed at the bottom of the machine, which can easily move the machine to a position and fix the casters afterwards.