E-mail
2715843753@qq.com
-
Phone
13077284859
-
Address
52 Datong Road, Jinshi City
Jinshi Petrochemical Instrument Co., Ltd
2715843753@qq.com
13077284859
52 Datong Road, Jinshi City
JSR3602Automatic micro residual carbon analyzer (micro method)
The automatic trace residual carbon analyzer is used to measure the residues formed by the evaporation and pyrolysis of petroleum products at high temperatures. To estimate the approximate trend of carbonaceous sediment formation under similar degradation conditions, and to provide an indicator of the relative coking tendency of petroleum products. Compliant with GB/T17144-1997 "Determination of residual carbon in petroleum products (micro method)" standard.
1、 Compliant withGB/T17144、ASTM D4530、ISO10370standard
2、 Summary of Experimental Methods
Place the weighed sample into a sample tube and heat it up to 500 ℃ in an inert gas (nitrogen) atmosphere according to the prescribed temperature program. The volatile substances generated during the reaction process are carried away by nitrogen gas, and the remaining carbonaceous residue accounts for the percentage of the original sample. Report the trace residual carbon value.
3、 Main technical parameters:
Heating temperature: room temperature~500±2℃
Temperature control: digital display temperature control temperature-programmed
Sample quantity:6~12Kong
Power supplyAC220V±10%50HZ
Overall dimensions:330×400×550=long×wide×High(mm)
Total weight of the machine23Kg
4、 Performance and features:
1Using a digital program temperature controller for temperature control, the temperature control accuracy is high, and the temperature display is intuitive and clear.
2Heat up to 500 ℃ at a rate of 10-15 ℃/min and automatically maintain a constant temperature. Then automatically cut off the heating power supply of the furnace to allow it to cool down naturally.
3It can automatically switch between two sets of nitrogen flow rates of 600mL/min and 150mL/min pre-set, and automatically control the blowing time without manual monitoring, which is very convenient.
4The connection method of the air inlet adopts a quick connector, which is easy to install and disassemble.
5、 Warranty and after-sales service:
The quality of the instruments produced by our company meets the standards and industry standards, with three guarantees for quality, a two-year warranty, and lifetime maintenance. Free installation and debugging, free transportation and packaging fees. The company will strictly follow the quality requirements undertaken by the project, carry out serious work, provide good technical services for all instruments, and accept user inquiries.
All prices are system default, please confirm with us for detailed prices!