The GPT-9600 series economical safety testing instrument is reasonably priced and can meet the routine testing of major safety regulations such as IEC, EN, UL, CSA, GB, JIS, and other safety rules.
The GPT-9600 series has the largest 100VA AC output power. GPT-9603 is a three in one model used for AC withstand voltage, DC withstand voltage, and insulation resistance testing; GPT-9612 is used for AC withstand voltage and insulation resistance testing; GPT-9602 is used for AC and DC withstand voltage testing; GPT-9601 is used for AC withstand voltage testing. Efficient PWM technology is its core, reducing the impact of AC input voltage fluctuations and providing a stable voltage for the GPT-9600 series.
| Specifications | ||
| AC withstand voltage | Output voltage range | 0.10kV~5.00kV ac |
| Output voltage resolution | 10V | |
| Zui Large Fixed Load | 100VA(5kV/20mA) | |
| Maximum rated current | 20mA(0.5kV<V≤5kV); 5mA(0.1kV≤V≤0.5kV) | |
| Output voltage waveform | sine wave | |
| Output voltage frequency | 50Hz/60Hz optional | |
| Voltage regulation rate | ± (1.5%+2 counts) [Full load → No load] | |
| Voltage meter accuracy | ± (1.5% reading+2 counts) | |
| Current measurement range | 0.01mA~20.0mA | |
| Current * Resolution | 0.01mA/0.1mA | |
| Accuracy of current measurement | ± (2.0% read value+5 counts) when HI SET<1.00mA; ± (2.0% reading+3 counts) when HI SET ≥ 1.00mA | |
| Accuracy of current determination | ± (3.0% set value+5 counts) when HI SET<1.00mA; ± (3.0% set value+3 counts) when HI SET ≥ 1.00mA | |
| Window comparison method | have | |
| ARC detection | have | |
| RAMP (Acceleration Time) | 0.1s fixed | |
| Timing (testing time) | OFF,1s~180s | |
| GND | ON | |
| DC withstand voltage | Output voltage range | 0.10kV~6.00kV dc |
| Output voltage resolution | 10V | |
| Zui Large Fixed Load | 25W(5kV/5mA) | |
| Maximum rated current | 6mA(0.5kV<V≤6kV); 2mA(0.1kV≤V≤0.5kV) | |
| Voltage regulation rate | ± (1.5%+2 counts) [Full load → No load] | |
| Voltage meter accuracy | ± (1.5% reading+2 counts) | |
| Current measurement range | 0.01mA~6.00mA | |
| Current * Resolution | 0.01mA | |
| Accuracy of current measurement | ± (2.0% read value+5 counts) when HI SET<1.00mA; ± (2.0% reading+3 counts) when HI SET ≥ 1.00mA | |
| Accuracy of current determination | ± (3.0% set value+5 counts) when HI SET<1.00mA; ± (3.0% set value+3 counts) when HI SET ≥ 1.00mA | |
| Window comparison method | have | |
| ARC detection | have | |
| RAMP (Acceleration Time) | 0.1s fixed | |
| Timing (testing time) | OFF,1s~180s | |
| GND | ON | |
| insulation resistance | output voltage | 50V, 100V, 250V, 500V, 1000V dc |
| Output voltage accuracy | ± (3.0% set value+1 count) | |
| Resistance measurement range | 1MΩ~2000MΩ | |
| test voltage | Measurement range accuracy | |
| 50V/100V/250V | 1~50M Ω± (5% reading+2M Ω) | |
| 51~2000M Ω± (10% reading+2M Ω) | ||
| 500V/1000V | 1~500M Ω± (5% reading+2M Ω) | |
| 501~2000M Ω± (10% reading+2M Ω) | ||
| Window comparison method | have | |
| output impedance | 600kΩ | |
| RAMP (Acceleration Time) | 0.1s fixed | |
| Timing (testing time) | OFF,1s~180s | |
| GND | OFF (fixed) | |
| Testing method* | single time | ACW, DCW, IR |
| automatic | AC-IR, IR-AC, DC-IR, IR-DC | |
| interface | Remote terminal (front) | standard configuration |
| Signal I/O | standard configuration | |
| display | 240 × 48 ice blue dot matrix LCD | |
| power supply | AC 100V~120V/220V~240V±10%,50/60Hz | |
| power loss | 400VA Max | |
| Size & Weight | 330(W)×148(H)×385(D)mm About 9kg max |
|
| *The 'test mode' is related to the selected model | ||