Display mode: LED (digital tube) display, high brightness, eye-catching and intuitive, with a long visible distance
Measurement range: electrode constant × (1-3000) μ S/cm, 0-99.9 ℃
0.01~30μS/cm ; (with 0.01 electrode);
0.1~300μS/cm ; (with 0.1 electrode);
1.0~3000μS/cm ; (Equipped with 1.0 electrode);
10-30000 μ S/cm (with 10.0 electrode);
Resolution: 0.001 μ S/cm, 0.1 ℃
Electrode constant: set by software between 0.001 and 15.0 cm-1
Temperature compensation range: automatic 0-99.9 ℃, manual 0-99.9 ℃, converted to 25 ℃
Accuracy: ± 0.5% FS, ± 0.3 ℃
Current isolation output: 4-20 mA (load<750 Ω).
Alarm relay: AC220V, 3A, alarm signal isolated output
Communication interface: isolated RS485
Power supply: AC220V ± 22V, 50Hz ± 1Hz
Protection level: IP65
Working conditions: ambient temperature 0-60 ℃, relative humidity ≤ 90 ℅
Power off storage:>10 years
Weight and size: 0.8kg, 146 × 146 × 108 mm
Hole size: 138x138mm;
Secondary meter installation methods: perforated/wall mounted/rack mounted/pipe mounted
Electrode installation methods: flow-through/submerged/flange/pipeline
Standard configuration: one secondary meter, one set of measuring cells, and one electrode