Single column electronic * testing machineCapable of conducting various mechanical tests at room temperature, the microcomputer control system automatically collects and processes test data, draws multiple curves, and prints test reports. The microcomputer control system software is based on the WINDOWS operating system platform and has the characteristics of fast running speed, friendly interface, and simple operation, which can meet the testing and measurement needs of different materials. The software can expand its functions according to user requirements, and the measurement and control software can expand its functions according to user requirements. It is widely used in petrochemical, mechanical manufacturing, plastic rubber, adhesives, ceramic building materials, metal materials, as well as higher education institutions, scientific research institutions, technical supervision, quality inspection stations and other departments.
1、 Host maintenance:
1. The fixtures equipped with the machine should be coated with rust proof oil for storage;
2. Due to frequent use of the jaws of the testing machine, they are prone to wear and tear, and excessive oxidation can easily cause damage to the small piston and oil leakage. Therefore, the jaws should be cleaned regularly to maintain cleanliness;
3. The sliding surface between the embedded steel plate and the lining plate, as well as the dovetail groove surface on the lining plate, should be kept clean and regularly coated with a thin layer of MoS2 (molybdenum disulfide) grease;
4. Regularly check the screws in the jaw area, and tighten them promptly if any looseness is found;
5. Regularly check the transmission condition of the sprocket. If any looseness is found, please re tighten the tensioning wheel;
2、 Maintenance of control system:
1. Regularly check whether the connecting wires on the back panel of the controller are in good contact, and if they are loose, they should be tightened in a timely manner;
2. If the machine is not used for a long period of time after the experiment, turn off the controller and computer;
3. The interfaces on the controller correspond one-to-one, and plugging in the wrong interface may cause damage to the device;
4. The interface on the controller must be powered off.
3、 Maintenance of oil source:
1. Regularly check whether there are any oil leaks at the host and oil source. If any oil leaks are found, the sealing ring or combination gasket should be replaced in a timely manner;
2. According toSingle column electronic * testing machineRegularly replace the oil suction filter and filter element, and replace the hydraulic oil according to the usage situation and oil usage period.
3. When not conducting experiments for a long time, pay attention to turning off the power supply of the host. If the machine is in standby mode, the transfer switch should be set to the "load" position, because if the transfer switch is set to the "fast reverse" position, the electromagnetic reversing valve will remain powered on, which will affect the service life of the device.