1、 No peak or very small peak:
1Possible reasons: The detector is turned off or not lit.
Exclusion method: Re ignite.
2Possible cause: Leakage at the injection port.
Exclusion method: Replace the rubber pad.
3Possible reason: The temperature in the gasification chamber is too low, and the sample cannot be gasified.
Exclusion method: Increase the gasification temperature.
4Possible reason: There is no air flow inside the column.
Exclusion method: Check.
5Possible reason: The connection between the chromatographic column and the injection port or detector is incorrect.
Exclusion method: Tighten the joint.
6Possible cause: Recorder malfunction.
Exclusion method: Maintenance.
7Possible cause: Blockage of detector nozzle.
Exclusion method: Cleaning.
2、 During isothermal conditions, the baseline drifts in one direction or irregularly:
1Possible cause: Leakage at the entrance of the column.
Exclusion method: Eliminate air leakage.
2Possible reason: The supplementary air flow rate is too low.
Exclusion method: Increase the tail air flow rate.
3Possible reason: The detector furnace temperature has changed and has not reached equilibrium.
Exclusion method: Allow sufficient time for the baseline to stabilize.
4Possible cause: Fixed liquid loses heat.
Exclusion method: Aging the chromatographic column and adjusting the temperature.
5Possible reasons:FIDThe flow rate of hydrogen gas varies.
Exclusion method: Check the pressure and flow rate control of the gas cylinder.
6Possible cause: Detector leakage.
Exclusion method: Replace the sealing gasket.
3、 Baseline increase during program heating:
1Possible cause: Exceeding the column temperature limit.
Exclusion method: Reduce high temperature.
2Possible cause: The column is contaminated.
Exclusion method: Check the gas filter and aging chromatographic column.
3Possible cause: Detector leakage.
Exclusion method: Replace the sealing gasket.
4Possible cause: Poor chromatographic column.
Exclusion method: Aging chromatographic column.
5Possible cause: Increased loss of column fixed phase.
Exclusion method: This phenomenon is normal within the temperature range of use.
4、 Poor precision and quantitative repeatability:
1Possible cause: Wear of the injection needle.
Exclusion method: Replace the injection needle.
2Possible reason: Poor injection technique.
Exclusion method: Improve the injection technique.
3Possible cause: contamination of the injection port.
Exclusion method: Clean the lining tube and replace the quartz wool.