Common troubleshooting methods for touch screen salt spray test chamber
Date: 2022-02-25Read: 19

Precision salt spray test chamberCommon troubleshooting methods
The use of salt spray test chambers in harsh environments may inevitably result in some minor malfunctions. It is time-consuming and labor-intensive to seek repairs from manufacturers, but some minor issues can be resolved by oneself. That's definitely a question. We don't know how to handle it technically. Don't worry, I'm here to share some common faults and solutions today so that everyone can easily solve the problem.
1. If the device cannot operate normally after turning on the power, you can consider checking if the water level in the heating water tank is too low. When the water level is too low, the device will cut off the power supply to ensure its safety; If the malfunction still cannot be operated normally after eliminating such faults, it must be a functional failure of the salt spray test chamber equipment. At this time, the manufacturer can be contacted for repair.
2. If the touch screen does not respond, we will check if the screen is automatically locked and unlock it; After eliminating the factor of screen lock, please contact the manufacturer for repair or replacement.
If the water level in the salt spray test chamber is too low, we can check whether the water level in the laboratory or saturation tank is sufficient.
4. If the temperature in the studio cannot rise to the set temperature, we can check if it is the temperature of the laboratory temperature controller that is set too low; If so, simply set the temperature controller to the desired temperature; 2) Check if the safety protection switch is set too low, just set the safety protection switch to the desired temperature; Huanrui Testing suggests that if the above two problems are ruled out and still cannot be solved, it may be a heating system malfunction, electromagnetic relay malfunction, or controller malfunction; Need to find the manufacturer for repair.
5. If the spray is not timed, we will immediately check whether the temperature rises to the temperature, because if the temperature does not reach the preset temperature, the timing equipment must think that the experiment has not yet started; Of course, it may also be that the set time is the same as the time of spray, so the timing time cannot be observed.
6. There is normal spray but the air compressor does not operate normally; In general, for the safety of the equipment, the air compressor of the salt spray test chamber is usually equipped with an automatic protection system. When the salt spray test chamber is used abnormally, the air compressor will automatically stop running. Therefore, when such a fault occurs, you can continue to use it as usual.
Customers can try to solve the problems mentioned above on their own, but it is important to be cautious. If you are unable to understand the underlying principles, you can first consult with the manufacturer of the salt spray test chamber to see if they can solve the problem. If not, you can have the manufacturer send technicians to come and repair the equipment. Do not disassemble the equipment without authorization to avoid causing larger malfunctions.